FDA Adverse Event Malfunction Summary report: N

CORE IMPACTION DRILL

MDR report key: 2121128 · Received May 4, 2011

Report

Report Number
1811755-2011-01581
Event Type
Malfunction
Date Received
May 4, 2011
Date of Event
April 11, 2011
Report Date
April 11, 2011
Manufacturer
STRYKER INSTRUMENTS KALAMAZOO
Product Code
ERL
PMA / PMN Number
K040300
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NY, US
Reporter Occupation
UNKNOWN

Narratives

Additional Manufacturer Narrative · 1

THE DEVICE WAS EVALUATED BY THE MANUFACTURER AND THE HANDPIECE EXCEEDED THE MAXIMUM ALLOWED TEMPERATURE RISE AND HAD EXCESSIVE NOISE COMING FROM THE MOTOR. THE CONTACT PINS WERE BURNT AND THERE WAS INSUFFICIENT LUBE IN THE BEARINGS ON THE ROTOR AND DRIVESHAFT ASSEMBLIES. THE DEVICE WAS REPAIRED AND RETURNED TO THE CUSTOMER.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE PRODUCT WAS HEATING UP DURING A WISDOM TOOTH REMOVAL. ANOTHER DRILL WAS PULLED TO COMPLETE THE PROCEDURE. THERE WAS NO MEDICAL INTERVENTION REQUIRED AND NO DELAY IN THE PROCEDURE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 CORE IMPACTION DRILL ERL STRYKER INSTRUMENTS KALAMAZOO

Patients

Seq Age Sex Outcome Treatment
1 UNK