FDA Adverse Event
Malfunction
Summary report: N
CORE IMPACTION DRILL
MDR report key: 2121128
·
Received May 4, 2011
Report
- Report Number
- 1811755-2011-01581
- Event Type
- Malfunction
- Date Received
- May 4, 2011
- Date of Event
- April 11, 2011
- Report Date
- April 11, 2011
- Manufacturer
- STRYKER INSTRUMENTS KALAMAZOO
- Product Code
- ERL
- PMA / PMN Number
- K040300
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NY, US
- Reporter Occupation
- UNKNOWN
Narratives
Additional Manufacturer Narrative · 1
THE DEVICE WAS EVALUATED BY THE MANUFACTURER AND THE HANDPIECE EXCEEDED THE MAXIMUM ALLOWED TEMPERATURE RISE AND HAD EXCESSIVE NOISE COMING FROM THE MOTOR. THE CONTACT PINS WERE BURNT AND THERE WAS INSUFFICIENT LUBE IN THE BEARINGS ON THE ROTOR AND DRIVESHAFT ASSEMBLIES. THE DEVICE WAS REPAIRED AND RETURNED TO THE CUSTOMER.
Description of Event or Problem · 1
IT WAS REPORTED THAT THE PRODUCT WAS HEATING UP DURING A WISDOM TOOTH REMOVAL. ANOTHER DRILL WAS PULLED TO COMPLETE THE PROCEDURE. THERE WAS NO MEDICAL INTERVENTION REQUIRED AND NO DELAY IN THE PROCEDURE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | CORE IMPACTION DRILL | ERL | STRYKER INSTRUMENTS KALAMAZOO |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | UNK |