FDA Adverse Event Injury Summary report: N

SPRINT FIDELIS

MDR report key: 2121124 · Received June 13, 2011

Report

Report Number
2649622-2011-07792
Event Type
Injury
Date Received
June 13, 2011
Report Date
April 23, 2020
Manufacturer
MEDTRONIC PUERTO RICO, INC.
Product Code
LWS
PMA / PMN Number
P920015
Removal / Correction Number
Z-0070-2008
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. DETAILED ANALYSIS OF THE DEVICE WAS NOT PERFORMED AT THIS TIME DUE TO PENDING LITIGATION. THEREFORE, WE ARE UNABLE TO FULLY EVALUATE THE REPORTED EVENT.

Additional Manufacturer Narrative · 1

PRODUCT EVENT SUMMARY: THE FULL LEAD WAS RETURNED IN SEGMENTS, ANALYZED, AND NO ANOMALIES WERE FOUND. VISUAL ANALYSIS OF THE LEAD INDICATED APPARENT EXPLANT DAMAGE. IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.

Description of Event or Problem · 1

THE PATIENT REPORTED EXPERIENCED "ELECTRICAL TINGLING" AND THE RIGHT VENTRICULAR LEAD HAD FRACTURED. IT WAS ALSO REPORTED THAT A LEAD INTEGRITY ALERT WAS HEARD DAILY. THE LEAD WAS EXPLANTED AND REPLACED. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Description of Event or Problem · 1

THE PATIENT REPORTED EXPERIENCED "ELECTRICAL TINGLING" AND THE RIGHT VENTRICULAR LEAD HAD FRACTURED. THE LEAD WAS EXPLANTED AND REPLACED. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 SPRINT FIDELIS DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER LWS MEDTRONIC PUERTO RICO, INC. 6949 ASKU

Patients

Seq Age Sex Outcome Treatment
1 31 YR Hospitalization| R 7288 IMPLANTABLE PACEMAKER/CARDIO/DEFIB| 7288 IMPLANTABLE PACEMAKER/CARDIO/DEFIB| 5076 IMPLANTABLE PACING LEAD| 5076 IMPLANTABLE PACING LEAD