ANIMAS INSULIN INFUSION PUMP
Report
- Report Number
- 2531779-2011-04079
- Event Type
- Malfunction
- Date Received
- June 13, 2011
- Report Date
- May 16, 2011
- Manufacturer
- ANIMAS CORPORATION
- Product Code
- LZG
- PMA / PMN Number
- K042873
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA
- Reporter Occupation
- OTHER
Narratives
(B)(4). DEVICE EVALUATION: THE PUMP HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON (B)(4) 2011 WITH THE FOLLOWING FINDINGS: EVALUATION REVEALED THAT THE OK KEYPAD BUTTON IS INTERMITTENTLY RESPONSIVE. THERE WAS EVIDENCE OF KEYPAD ADHESIVE FOUND UNDER ALL KEY CONTACTS. THE KEYPAD WAS FOUND TO BE LIFTING OVER THE UP ARROW KEYPAD BUTTON. A DAMAGED KEYPAD WILL PERMIT CONTAMINATION TO PERMEATE THE BUTTONS WHICH WILL HAVE A NEGATIVE IMPACT ON BUTTON FUNCTION. THIS SITUATION IS NOT LIKELY TO RESULT IN AN ADVERSE EVENT AS THE DAMAGED KEYPAD SHOULD BE CLEARLY VISIBLE AND WARNS THE PATIENT TO DISCONTINUE USING THE PUMP. THE OWNER'S BOOKLET INSTRUCTS THE USER TO CALL ANIMAS CUSTOMER SERVICE IF THE USER SUSPECTS THAT THE PRODUCT IS DAMAGED. ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE.
THE PUMP HAS NOT BEEN RETURNED TO ANIMAS FOR EVALUATION. IF THE DEVICE IS RETURNED, AN EVALUATION SHALL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSIONS CAN BE MADE AT THIS TIME.
PATIENT REPORTS THAT OK BUTTON IS NOT ALWAYS RESPONDING FOR THE LAST COUPLE WEEKS. PATIENT DOES NOT CLEAN THE PUMP. PATIENT WEARS THE PUMP IN UNDER GARMENT AGAINST BODY.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ANIMAS INSULIN INFUSION PUMP | INSULIN INFUSION PUMP | LZG | ANIMAS CORPORATION | ANM 2020 INSULIN INFUSION PUMP |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 17 YR |