FDA Adverse Event Malfunction Summary report: N

ZERO TIP NITINOL STONE RETRIEVAL BASKET

MDR report key: 2121094 · Received June 13, 2011

Report

Report Number
3005099803-2011-01910
Event Type
Malfunction
Date Received
June 13, 2011
Report Date
May 17, 2011
Manufacturer
BOSTON SCIENTIFIC - SPENCER
Product Code
FFL
PMA / PMN Number
EXEMPT
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

ANALYSIS OF THE RETURNED ZERO TIP NITINOL RETRIEVAL BASKET REVEALED THE BASKET AND PULL WIRE REMAINED ATTACHED TO THE HANDLE. ALL OF THE BASKET WIRES WERE PRESENT AND ATTACHED. THE STRAIN RELIEFS (BLACK AND BLUE HEAT SHRINKS) AND OUTER SHEATH WERE DETACHED FROM THE FEMALE LUER AT THE DISTAL END OF THE HANDLE. THE FEMALE LUER COMPONENT ((B)(6)) WAS BROKEN. THE PULL WIRE WAS KINKED AT THE DISTAL END OF THE HANDLE. KINKS WERE FOUND ALONG THE OUTER SHEATH WORKING LENGTH. WITH THE LUER HELD INTO PLACE ON THE HANDLE WHILE ACTUATING, THE BASKET WOULD OPEN AND CLOSE SEVERAL TIMES WITHOUT ISSUE WHILE HELD STRAIGHT AND WHEN HELD IN A LOOP. THEREFORE, THE MOST PROBABLE ROOT CAUSE FOR THIS COMPLAINT IS OPERATIONAL/PHYSIOLOGICAL CONTEXT. IT SHOULD ALSO BE NOTED THAT THE SHEATH DETACHING FROM THE HANDLE IS NOT A MEDWATCH REPORTABLE EVENT.

Description of Event or Problem · 1

IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A ZERO TIP NITINOL SONE RETRIEVAL BASKET WAS USED IN A FLEXIBLE URETEROSCOPY AND LASER LITHOTRIPSY PROCEDURE TO TREAT A PATIENT WITH RENAL STONES (PATIENT IDENTIFIER, AGE, SEX, AND WEIGHT UNKNOWN). ACCORDING TO THE COMPLAINANT, THE BASKET BROKE OFF RIGHT AT THE BASE OF THE HANDLE. NO ADDITIONAL INFORMATION IS AVAILABLE AT THIS TIME.

Description of Event or Problem · 1

IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A ZERO TIP NITINOL SONE RETRIEVAL BASKET WAS USED IN A FLEXIBLE URETEROSCOPY AND LASER LITHOTRIPSY PROCEDURE TO TREAT A PATIENT WITH RENAL STONES (PATIENT IDENTIFIER, AGE, SEX, AND WEIGHT UNKNOWN). ACCORDING TO THE COMPLAINANT, THE BASKET BROKE OFF RIGHT AT THE BASE OF THE HANDLE. NO ADDITIONAL INFORMATION IS AVAILABLE AT THIS TIME.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ZERO TIP NITINOL STONE RETRIEVAL BASKET DISLODGER, STONE, BASKET, URETERAL, METAL FFL BOSTON SCIENTIFIC - SPENCER M0063901050 14197765

Patients

Seq Age Sex Outcome Treatment
1