FDA Adverse Event Malfunction Summary report: N

COLLEAGUE PRE P1.7

MDR report key: 2120998 · Received June 13, 2011

Report

Report Number
6000001-2011-07566
Event Type
Malfunction
Date Received
June 13, 2011
Date of Event
May 31, 2011
Report Date
June 1, 2011
Manufacturer
BAXTER HEALTHCARE - SINGAPORE
Product Code
FRN
PMA / PMN Number
K063696
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
OH, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). EVALUATION SUMMARY: THE REPORTED CONDITION OF A COLLEAGUE INFUSION PUMP WITH DAMAGED BATTERY WAS CONFIRMED DURING PRODUCT EVALUATION. THE ASSIGNABLE CAUSE FOR THE REPORTED PROBLEM WAS DETERMINED TO BE DEPLETED BATTERIES RESULTING FROM USER ERROR. APPROPRIATE ACTION WAS TAKEN TO CORRECT THE REPORTED PROBLEM BY REPLACING THE MAIN BATTERIES AND HARNESS. THE DEVICE HAS BEEN PREVIOUSLY SENT INTO SERVICE FOR THE REPORTED CONDITION OF "DAMAGED BATTERY." DURING PREVIOUS SERVICES, THE MAIN BATTERIES AND HARNESS WERE REPLACED. THIS IS INVOLVING A PUMP WITH SOFTWARE VERSION 6.13.90 WHICH IS CATEGORIZED AS A COLLEAGUE 2006.

Additional Manufacturer Narrative · 1

(B)(4). THE DEVICE IS CURRENTLY IN THE PROCESS OF BEING EVALUATED ON SITE BY A BAXTER FIELD SERVICE TECHNICIAN. A FOLLOW-UP REPORT WILL BE FILED UPON COMPLETION OF THE EVALUATION OR IF ANY ADDITIONAL DETAILS BECOME AVAILABLE.

Additional Manufacturer Narrative · 1

(B)(4).

Description of Event or Problem · 1

THE FACILITY REPORTED A COLLEAGUE INFUSION PUMP WITH A MALFUNCTION OF DAMAGED BATTERY. THIS CONDITION HAD THE POTENTIAL TO INTERRUPT DELIVERY. IT IS UNKNOWN WHEN THIS CONDITION OCCURRED. ACCORDING TO THE HOSPITAL REPRESENTATIVE, THERE WAS NO PATIENT INVOLVEMENT. NO PATIENT INJURY OR MEDICAL INTERVENTION HAD BEEN REPORTED RELATED TO THE DEVICE. NO ADDITIONAL INFORMATION IS AVAILABLE. THE USER INTERFACE MODULE SOFTWARE VERSION OF THIS DEVICE IS CURRENTLY UNKNOWN.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 COLLEAGUE PRE P1.7 PUMP, INFUSION FRN BAXTER HEALTHCARE - SINGAPORE

Patients

Seq Age Sex Outcome Treatment
1