FDA Adverse Event Malfunction Summary report: N

HOMECHOICE

MDR report key: 2120994 · Received June 13, 2011

Report

Report Number
1423500-2011-07500
Event Type
Malfunction
Date Received
June 13, 2011
Date of Event
March 20, 2011
Report Date
May 25, 2011
Manufacturer
BAXTER HEALTHCARE - SINGAPORE
Product Code
FKX
PMA / PMN Number
K102936
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE DEVICE HAS BEEN RECEIVED, AND THE EVALUATION IS IN PROCESS. A FOLLOW-UP MDR WILL BE SUBMITTED UPON COMPLETION OF THE EVALUATION OR IF ANY ADDITIONAL INFORMATION IS RECEIVED.

Additional Manufacturer Narrative · 1

(B)(4). EVALUATION SUMMARY: THE PAL (PRODUCT ANALYSIS LAB) EVALUATED THE DEVICE. A REVIEW OF THE DEVICE LOGS REVEALED A AN INCREASED INTRA-PERITONEAL VOLUME (IIPV). THE CAUSE WAS DETERMINED TO BE INSUFFICIENT DRAIN, USE ERROR AS THE TIDAL ULTRAFILTRATION REMOVAL WAS SET TOO LOW. THE DEVICE HISTORY RECORD WAS REVIEWED AND NO ISSUES WERE IDENTIFIED THAT MAY HAVE CONTRIBUTED TO THE IIPV. A SERVICE HISTORY REVIEW REVEALED NO PREVIOUS SERVICE EVENTS WERE RELATED TO THE IIPV. LABELING REVIEW FOUND LABELING TO BE ADEQUATE FOR THE USER ERROR IDENTIFIED IN THIS REPORT. BAXTER HAS CONDUCTED A TREND REVIEW AND FOUND THAT SIMILAR REPORTS HAVE BEEN RECEIVED FOR THE REPORTED PROBLEM. THIS ISSUE HAS BEEN ESCALATED TO CAPA.

Description of Event or Problem · 1

DURING INITIAL ASSESSMENT OF A RETURNED HOMECHOICE MACHINE, A BAXTER TECHNICIAN FOUND AN INCREASED INTRA-PERITONEAL VOLUME (IIPV) THAT WAS IDENTIFIED IN PATIENT THERAPY LOG ON (B)(6) 2011 AT 22:03 WITH ULTRAFILTRATION OF 2174 ML DURING CYCLE 4. THERE WAS PATIENT INVOLVEMENT BUT NO PATIENT INJURY OR MEDICAL INTERVENTION INDICATED. THIS EVENT MEETS OVERFILL CRITERIA.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 HOMECHOICE SYSTEM, PERITONEAL, AUTOMATIC DELIVERY FKX BAXTER HEALTHCARE - SINGAPORE

Patients

Seq Age Sex Outcome Treatment
1