FDA Adverse Event Malfunction Summary report: N

STERLING BALLOON DILATATION CATHETER

MDR report key: 2120983 · Received June 13, 2011

Report

Report Number
2134265-2011-02420
Event Type
Malfunction
Date Received
June 13, 2011
Date of Event
May 31, 2011
Report Date
May 31, 2011
Manufacturer
BOSTON SCIENTIFIC - MAPLE GROVE
Product Code
DQY
PMA / PMN Number
K053116
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

AGE AT TIME OF EVENT: 18 YEARS OR OLDER. DEVICE EVALUATED BY MFR.: THE COMPLAINT DEVICE WAS NOT RECEIVED FOR ANALYSIS. THE MANUFACTURING BATCH RECORD REVIEW CONFIRMED THAT THE DEVICE MET ALL MATERIAL, ASSEMBLY AND PERFORMANCE SPECIFICATIONS. THE MOST PROBABLE ROOT CAUSE IS CONSIDERED HANDLING DAMAGE AS THE EVENT OCCURRED PRIOR TO PATIENT CONTACT. (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING PREPARATION FOR A PERCUTANEOUS TRANSLUMINAL ANGIOPLASTY (PTA) TREATMENT PROCEDURE, A BALLOON RUPTURE OCCURRED. THE 90% STENOSED LESION WAS LOCATED IN A NON CALCIFIED AND MODERATELY TORTUOUS VEIN SHUNT IN THE ARM. UPON REMOVING AIR FROM THE 5.0MM X 40MM STERLING BALLOON CATHETER DURING PREP, THE BALLOON RUPTURED. THE PROCEDURE WAS COMPLETED WITH A DIFFERENT DEVICE. NO PATIENT COMPLICATIONS WERE REPORTED AND THE PATIENT'S STATUS IS GOOD.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 STERLING BALLOON DILATATION CATHETER CATHETER, PERCUTANEOUS DQY BOSTON SCIENTIFIC - MAPLE GROVE H74939032504080 14239027

Patients

Seq Age Sex Outcome Treatment
1