FDA Adverse Event Malfunction Summary report: N

INFANT CONTINUOUS FLOW BREATHING CIRCUIT

MDR report key: 2120912 · Received June 12, 2011

Report

Report Number
9611451-2011-00349
Event Type
Malfunction
Date Received
June 12, 2011
Date of Event
May 13, 2011
Report Date
May 16, 2011
Manufacturer
FISHER & PAYKEL HEALTHCARE LIMITED
Product Code
BTT
PMA / PMN Number
K020332
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE COMPLAINT RT329 INFANT CONTINUOUS FLOW BREATHING CIRCUIT KIT IS EN ROUTE TO FISHER & PAYKEL HEALTHCARE FOR INVESTIGATION. WE WILL PROVIDE A FOLLOW-UP REPORT ONCE WE RECEIVE THE COMPLAINT DEVICE AND HAVE COMPLETED OUR INVESTIGATION.

Additional Manufacturer Narrative · 1

(B)(4). METHOD: THE RT329 INFANT CONTINUOUS FLOW BREATHING CIRCUIT KIT, WHICH WAS RETURNED UNSEALED, WAS VISUALLY INSPECTED FOR A MISSING ADAPTOR. RESULTS: A VISUAL INSPECTION REVEALED THAT THE OFFSET ADAPTOR WAS MISSING FROM THE BREATHING CIRCUIT KIT, CONFIRMING THE REPORTED FAULT. A LOT CHECK REVEALED NO OTHER COMPLAINTS OF THIS NATURE FOR LOT NUMBER 101208. CONCLUSION: THE BREATHING CIRCUIT PACKAGE CONSISTS OF A NUMBER OF COMPONENTS GROUPED TOGETHER DURING PRODUCTION. IT IS LIKELY THAT OPERATOR ERROR, DURING THE PACKING PROCESS, RESULTED IN THE MISSING OFFSET ADAPTOR. (B)(4). BREATHING CIRCUIT KITS ARE ALSO VISUALLY INSPECTED FOR MISSING COMPONENTS PRIOR TO DISTRIBUTION. (B)(4).

Description of Event or Problem · 1

A HEALTHCARE FACILITY IN (B)(6) REPORTED TO A FISHER & PAYKEL HEALTHCARE (FPH) FIELD REPRESENTATIVE THAT AN RT064 ADAPTER WAS MISSING FROM AN RT329 INFANT CONTINUOUS FLOW BREATHING CIRCUIT KIT. THIS WAS NOTICED PRIOR TO PATIENT USE. NO PATIENT CONSEQUENCE WAS REPORTED.

Description of Event or Problem · 1

A HEALTHCARE FACILITY IN (B)(6) REPORTED TO A FISHER & PAYKEL HEALTHCARE (FPH) FIELD REPRESENTATIVE THAT AN RT064 ADAPTER WAS MISSING FROM AN RT329 INFANT CONTINUOUS FLOW BREATHING CIRCUIT KIT. THIS WAS NOTICED PRIOR TO PATIENT USE. NO PATIENT CONSEQUENCE WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 INFANT CONTINUOUS FLOW BREATHING CIRCUIT BTT BTT FISHER & PAYKEL HEALTHCARE LIMITED RT329 101208