COLLEAGUE PRE P1.7
Report
- Report Number
- 6000001-2011-07545
- Event Type
- Malfunction
- Date Received
- June 11, 2011
- Date of Event
- May 30, 2011
- Report Date
- May 31, 2011
- Manufacturer
- BAXTER HEALTHCARE - SINGAPORE
- Product Code
- FRN
- PMA / PMN Number
- K063696
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- KY, US
- Reporter Occupation
- OTHER
Narratives
(B)(4). THE DEVICE WAS RETURNED TO BAXTER AND IS CURRENTLY IN THE PROCESS OF BEING EVALUATED. A FOLLOW-UP REPORT WILL BE FILED UPON COMPLETION OF THE EVALUATION OR IF ANY ADDITIONAL DETAILS BECOME AVAILABLE.
(B)(4). EVALUATION SUMMARY: THE REPORTED CONDITION OF A COLLEAGUE INFUSION PUMP WITH A LOW LITHIUM BATTERY WAS CONFIRMED AND DUPLICATED DURING PRODUCT EVALUATION. THE ROOT CAUSE WAS ASSIGNED TO LOW/DEPLETED LITHIUM BATTERY. THE LITHIUM BATTERY WAS REPLACED IN ORDER TO CORRECT THIS CONDITION. A SERVICE HISTORY REVIEW REVEALED NO PREVIOUS SERVICE EVENTS WERE RELATED TO THE REPORTED CONDITION. THIS IS INVOLVING A PUMP WITH SOFTWARE VERSION 5.09.90 WHICH IS CATEGORIZED AS A COLLEAGUE 2006.
(B)(4). INITIAL DATE RECEIVED BY MANUFACTURER WAS REPORTED INCORRECTLY; IT SHOULD HAVE BEEN (B)(6) 2011.
THE FACILITY REPRESENTATIVE CONTACTED A TECHNICAL SERVICE REPRESENTATIVE TO REPORT A COLLEAGUE INFUSION PUMP WITH THE REPORTED CONDITION "LOW LITHIUM BATTERY." THIS CONDITION WAS FOUND IN THE ER DEPARTMENT. THIS EVENT OCCURRED DURING PROGRAMMING/SETUP. THE FACILITY REPRESENTATIVE STATED THAT THERE WAS NO KNOWN PATIENT INVOLVEMENT AND NO REPORTS OF PATIENT INJURY OR MEDICAL INTERVENTION. NO ADDITIONAL INFORMATION IS AVAILABLE.