HOMECHOICE
Report
- Report Number
- 1423500-2011-07472
- Event Type
- Malfunction
- Date Received
- June 11, 2011
- Date of Event
- January 8, 2011
- Report Date
- January 27, 2011
- Manufacturer
- BAXTER HEALTHCARE - LARGO
- Product Code
- FKX
- PMA / PMN Number
- K053512
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FL, US
- Reporter Occupation
- OTHER
Narratives
(B)(4). THE DEVICE HAS BEEN RECEIVED BY BAXTER, AND THE EVALUATION HAS NOT YET BEEN COMPLETED. A FOLLOW-UP MDR WILL BE SUBMITTED UPON COMPLETION OF THE EVALUATION OR IF ANY ADDITIONAL INFORMATION IS RECEIVED.
(B)(4). THE DEVICE WAS RECEIVED AND EVALUATED. AN INSTANCE OF IIPV WAS REVEALED IN THE LOGS. ROOT CAUSE: INSUFFICIENT DRAIN - FALSE EMPTY DETECT AT INITIAL DRAIN AND AT PREVIOUS THERAPY LAST DRAIN. REVIEW OF THE SERVICE DATES AND ACTIVITIES REVEALED THE PREVIOUS RETURN OF THE DEVICE WAS NOT FOR THE REPORTED PROBLEM OF IIPV. THE DEVICE MET SPECIFICATIONS. BAXTER HAS CONDUCTED A TREND REVIEW AND FOUND THAT SIMILAR REPORTS HAVE BEEN RECEIVED FOR THE REPORTED PROBLEM. THE ROOT CAUSE INVESTIGATION IS IN PROGRESS THROUGH (B)(4).
AN INCREASED INTRAPERITONEAL VOLUME (IIPV) SITUATION WAS IDENTIFIED IN THE LOG OF A RETURNED HOMECHOICE DEVICE. THE IIPV OCCURRED ON (B)(6) 2011, DURING DRAIN CYCLE3. THE TOTAL DRAIN VOLUME WAS 3342ML. THIS DRAIN VOLUME MEETS BAXTER'S IIPV CRITERIA. THERE WAS PATIENT INVOLVEMENT, BUT NO PATIENT INJURY OR MEDICAL INTERVENTION WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | HOMECHOICE | SYSTEM, PERITONEAL, AUTOMATIC DELIVERY | FKX | BAXTER HEALTHCARE - LARGO |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 76 YR | HOME CHOICE |