FDA Adverse Event Malfunction Summary report: N

HOMECHOICE

MDR report key: 2120862 · Received June 11, 2011

Report

Report Number
1423500-2011-07472
Event Type
Malfunction
Date Received
June 11, 2011
Date of Event
January 8, 2011
Report Date
January 27, 2011
Manufacturer
BAXTER HEALTHCARE - LARGO
Product Code
FKX
PMA / PMN Number
K053512
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE DEVICE HAS BEEN RECEIVED BY BAXTER, AND THE EVALUATION HAS NOT YET BEEN COMPLETED. A FOLLOW-UP MDR WILL BE SUBMITTED UPON COMPLETION OF THE EVALUATION OR IF ANY ADDITIONAL INFORMATION IS RECEIVED.

Additional Manufacturer Narrative · 1

(B)(4). THE DEVICE WAS RECEIVED AND EVALUATED. AN INSTANCE OF IIPV WAS REVEALED IN THE LOGS. ROOT CAUSE: INSUFFICIENT DRAIN - FALSE EMPTY DETECT AT INITIAL DRAIN AND AT PREVIOUS THERAPY LAST DRAIN. REVIEW OF THE SERVICE DATES AND ACTIVITIES REVEALED THE PREVIOUS RETURN OF THE DEVICE WAS NOT FOR THE REPORTED PROBLEM OF IIPV. THE DEVICE MET SPECIFICATIONS. BAXTER HAS CONDUCTED A TREND REVIEW AND FOUND THAT SIMILAR REPORTS HAVE BEEN RECEIVED FOR THE REPORTED PROBLEM. THE ROOT CAUSE INVESTIGATION IS IN PROGRESS THROUGH (B)(4).

Description of Event or Problem · 1

AN INCREASED INTRAPERITONEAL VOLUME (IIPV) SITUATION WAS IDENTIFIED IN THE LOG OF A RETURNED HOMECHOICE DEVICE. THE IIPV OCCURRED ON (B)(6) 2011, DURING DRAIN CYCLE3. THE TOTAL DRAIN VOLUME WAS 3342ML. THIS DRAIN VOLUME MEETS BAXTER'S IIPV CRITERIA. THERE WAS PATIENT INVOLVEMENT, BUT NO PATIENT INJURY OR MEDICAL INTERVENTION WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 HOMECHOICE SYSTEM, PERITONEAL, AUTOMATIC DELIVERY FKX BAXTER HEALTHCARE - LARGO

Patients

Seq Age Sex Outcome Treatment
1 76 YR HOME CHOICE