COLLEAGUE PRE P1.7
Report
- Report Number
- 6000001-2011-07522
- Event Type
- Malfunction
- Date Received
- June 10, 2011
- Date of Event
- June 1, 2011
- Report Date
- June 1, 2011
- Manufacturer
- BAXTER HEALTHCARE - SINGAPORE
- Product Code
- FRN
- PMA / PMN Number
- K063696
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NJ, US
- Reporter Occupation
- OTHER
Narratives
(B)(4). THE DEVICE WAS RETURNED TO BAXTER AND IS CURRENTLY AWAITING EVALUATION. A FOLLOW-UP REPORT WILL BE FILED UPON COMPLETION OF THE EVALUATION OR IF ANY ADDITIONAL DETAILS BECOME AVAILABLE.
(B)(4).
(B)(4). EVALUATION SUMMARY: THE REPORTED CONDITION OF A COLLEAGUE INFUSION PUMP WITH A DAMAGED BATTERY WAS CONFIRMED AND DUPLICATED DURING PRODUCT EVALUATION. THE ROOT CAUSE OF THIS CONDITION WAS ASSIGNED TO DEPLETED MAIN BATTERIES RESULTING FROM USER ERROR. THE MAIN BATTERIES AND BATTERY HARNESS WERE REPLACED TO CORRECT THIS CONDITION. THIS IS INVOLVING A PUMP WITH SOFTWARE VERSION 5.09.90 WHICH IS CATEGORIZED AS A COLLEAGUE 2006.
THE FACILITY REPORTED A COLLEAGUE INFUSION PUMP WITH A MALFUNCTION OF DAMAGED BATTERY. THIS CONDITION HAD THE POTENTIAL TO INTERRUPT DELIVERY. IT IS UNKNOWN WHEN THIS CONDITION OCCURRED. ACCORDING TO THE HOSPITAL REPRESENTATIVE, THERE WAS NO PATIENT INVOLVEMENT. NO PATIENT INJURY OR MEDICAL INTERVENTION HAD BEEN REPORTED RELATED TO THE DEVICE. NO ADDITIONAL INFORMATION IS AVAILABLE. THE SOFTWARE VERSION IS CURRENTLY UNKNOWN.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | COLLEAGUE PRE P1.7 | PUMP, INFUSION | FRN | BAXTER HEALTHCARE - SINGAPORE |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |