FDA Adverse Event Malfunction Summary report: N

COATED VICRYL PLUS ANTIBACTERIAL (POLYGLACTIN 910)

MDR report key: 2120589 · Received June 10, 2011

Report

Report Number
2210968-2011-00756
Event Type
Malfunction
Date Received
June 10, 2011
Date of Event
May 7, 2011
Report Date
May 17, 2011
Manufacturer
ETHICON, INC.
Product Code
GAM
PMA / PMN Number
K032420
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
HK
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

DATE SENT TO THE FDA: (B)(6), 2011. CONCLUSION: NO CONCLUSION CAN BE DRAWN AT THIS TIME. SHOULD ADDITIONAL INFORMATION BE OBTAINED, A SUPPLEMENTAL 3500A FORM WILL BE SUBMITTED ACCORDINGLY. IN ADDITION, A REVIEW OF THE BATCH MANUFACTURING RECORDS WAS CONDUCTED AND THE BATCH MET ALL FINISHED GOODS RELEASE CRITERIA.

Additional Manufacturer Narrative · 1

(B)(4). CONCLUSION: THE ACTUAL NEEDLE REVEALED INDENTS AND SCUFF MARKS AROUND THE BREAK AREA PRODUCED DURING HANDLING BY THE SURGICAL NEEDLE HOLDERS OR SOME OTHER GRIPPING DEVICE. THE NEEDLE FRACTURED DUE TO TENSILE OVERLOAD GENERATED DURING SEVERE MECHANICAL DEFORMATION, WITH SIGNS OF DUCTILE FAILURE. THERE ARE NO SIGNS OF MANUFACTURING DEFECTS FOUND ON THE NEEDLE. A VISUAL INSPECTION WAS ALSO PERFORMED ON ONE ADDITIONAL LOOSE NEEDLES RETURNED FOR EVALUATION. THERE WERE NO SIGNS OF MANUFACTURING DEFECTS FOUND ON THE NEEDLE. THERE WERE SCUFF MARKS AND INDENTS ON THE NEEDLE THAT APPEAR TO HAVE COME FROM HANDLING BY THE SURGICAL NEEDLE HOLDERS OR SOME OTHER GRIPPING DEVICE.

Description of Event or Problem · 1

IT WAS REPORTED THAT A PATIENT UNDERWENT A TOTAL ABDOMINAL HYSTERECTOMY PROCEDURE ON (B)(6), 2011 AND SUTURE WAS USED. DURING SUTURING OF THE CERVIX, THE NEEDLE TIP WAS MISSING. THE PATIENT WAS THEN EXAMINED UNDER X- RAY TO FIND THE MISSING NEEDLE TIP BUT NOTHING WAS FOUND. THERE WERE NO ADVERSE PATIENT CONSEQUENCES.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 COATED VICRYL PLUS ANTIBACTERIAL (POLYGLACTIN 910) SUTURE, ABSORBABLE GAM ETHICON, INC. NA CC2222

Patients

Seq Age Sex Outcome Treatment
1 44 YR