FDA Adverse Event Malfunction Summary report: N

ENDOPATH** XCEL* TROCAR

MDR report key: 2120502 · Received June 10, 2011

Report

Report Number
3005075853-2011-02359
Event Type
Malfunction
Date Received
June 10, 2011
Report Date
May 18, 2011
Manufacturer
ETHICON ENDO-SURGERY, LLC.
Product Code
GCJ
PMA / PMN Number
K032676
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FR
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE ANALYSIS RESULTS FOUND THAT THE DEVICE WAS RETURNED IN GOOD VISUAL CONDITION. UPON EVALUATION OF THE DEVICE, THE DUCKBILL WAS FOUND TO BE SLIGHTLY OPEN AT SLIT. A LEAK TEST WAS PERFORMED AND THE DEVICE WAS NOTED TO LEAK DURING INSERTION AND REMOVAL OF THE TEST PROBE THROUGH THE DEVICE. AS EACH DEVICE IS VISUALLY INSPECTED AND FUNCTIONALLY TESTED DURING THE MANUFACTURING PROCESS, NO CONCLUSION COULD BE REACHED AS TO WHAT MIGHT HAVE CAUSED THE REPORTED EVENT. HOWEVER, AN INVESTIGATION HAS BEEN INITIATED TO ADDRESS THE POTENTIAL ROOT CAUSE OF THE INSUFFLATIONS ISSUES. THE BATCH RECORD WAS REVIEWED AND NO ANOMALIES WERE NOTED DURING THE MANUFACTURING PROCESS.

Additional Manufacturer Narrative · 1

(B)(4). INFORMATION ANTICIPATED, BUT UNAVAILABLE AT THIS TIME.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING AN UNKNOWN PROCEDURE, THE DEVICE WOULD NOT INSUFFLATE. SINCE THE BEGINNING OF THE SURGERY, THE SURGEON FOUND A LEAK IN THE TROCAR. IT WAS IMPOSSIBLE TO BLOW UP THE ABDOMEN OF THE PATIENT. THE SURGEON USED A SECOND TROCAR TO PERFORM THE PROCEDURE. THERE WERE NO ADVERSE CONSEQUENCES FOR THE PATIENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ENDOPATH** XCEL* TROCAR LAPAROSCOPE, GENERAL AND PLASTIC SURGERY GCJ ETHICON ENDO-SURGERY, LLC. NA UNK

Patients

Seq Age Sex Outcome Treatment
1