FDA Adverse Event Other Summary report: N

POWERED WHEELCHAIR

MDR report key: 2120467 · Received June 3, 2011

Report

Report Number
1525712-2011-00236
Event Type
Other
Date Received
June 3, 2011
Date of Event
April 19, 2011
Report Date
February 20, 2012
Manufacturer
INVACARE
Product Code
ITI
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
KY, US
Reporter Occupation
MEDICAL EQUIPMENT COMPANY TECHNICIAN/REPRESENTATIVE

Narratives

Additional Manufacturer Narrative · 1

NO SERIOUS INJURY ALLEGED. CONSUMER STATED THAT SHE HEARD A POPPING NOISE AND SMELLED SMOKE. CHAIR WAS APPROXIMATELY ONE YEAR OLD AT TIME OF INCIDENT. PRODUCT HAS NOT BEEN RETURNED FOR EVALUATION AT THIS TIME SO IT IS UNKNOWN IF A MALFUNCTION OCCURRED OR IF OTHER FACTORS SUCH AS MISUSE, ABUSE, OR LACK OF MAINTENANCE MAY HAVE CAUSED OR CONTRIBUTED TO THIS INCIDENT. MDR FILED BASED ON ALLEGED UNCONFIRMED MALFUNCTION.

Additional Manufacturer Narrative · 1

INITIAL (B)(4) ISSUED MFG REPORT #1525712-2011-00236. INTERNAL INSPECTION REVEALED HEAT DAMAGE. THE OVERHEATING OF THE PCB AND COMPONENTS IN THE AREA OF THE OUTPUT FETS OCCURRED. THIS FAILURE IS COVERED UNDER RISK ASSESSMENT 119. DESIGN IS VIEWED AS ALARP. SOME DEGREE OF SMOKING LIKELY OCCURRED; HOWEVER, THIS WAS NOT AN INDICATION OF SUSTAINED COMBUSTION. THE ELECTRONIC MALFUNCTION OF THE CONTROLLER, INCLUDING ANY DEBRIS THAT RESULTED, WAS CONTAINED WITHIN THE METAL ENCLOSURE OF THE DEVICE. THE CABLES SHOWED NO SIGNS OF OVERHEATING. NO RISK OF SHOCK WAS PRESENT TO THE USER. NO SERIOUS INJURY ALLEGED. CONSUMER STATED THAT SHE HEARD A POPPING NOISE AND SMELLED SMOKE. CHAIR WAS APPROXIMATELY ONE YEAR OLD AT TIME OF INCIDENT. PRODUCT HAS NOT BEEN RETURNED FOR EVALUATION AT THIS TIME SO IT IS UNKNOWN IF A MALFUNCTION OCCURRED OR IF OTHER FACTORS SUCH AS MISUSE, ABUSE, OR LACK OF MAINTENANCE MAY HAVE CAUSED OR CONTRIBUTED TO THIS INCIDENT. MDR FILED BASED ON ALLEGED UNCONFIRMED MALFUNCTION.

Description of Event or Problem · 1

THE CONSUMER ALLEGES THE TRAM MODULE IS "FRIED." THE MODULE ALLEGEDLY MADE A POPPING NOISE AND SMELLS LIKE SMOKE. NO INJURY IS ALLEGED.

Description of Event or Problem · 1

THE CONSUMER ALLEGES THE TRAM MODULE IS "FRIED." THE MODULE ALLEGEDLY MADE A POPPING NOISE AND SMELLS LIKE SMOKE. NO INJURY IS ALLEGED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 POWERED WHEELCHAIR 890.3860 ITI INVACARE 3GTQ-MCG

Patients

Seq Age Sex Outcome Treatment
1 Other