FDA Adverse Event Injury Summary report: N

MICRO-TECH

MDR report key: 2120458 · Received May 9, 2011

Report

Report Number
1929691-2011-00004
Event Type
Injury
Date Received
May 9, 2011
Date of Event
December 17, 2010
Report Date
May 26, 2011
Manufacturer
STANLEY SECURITY SOLUTIONS, INC.
Product Code
KMI
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

ON (B)(6) 2011, (B)(6) RECEIVED THE EQUIPMENT AS RETURNED BY THE COMPLAINANT. DURING (B)(6) TEST, THE FOLLOWING WAS FOUND: THE MICRO-TECH INFORMER MATE MODEL # 81781 WAS WORKING ACCORDING TO THE MANUFACTURER SPECIFICATIONS. THE MICRO-TECH PAD MODEL #82710 (LOT # 30110) WAS WORKING ACCORDING TO MANUFACTURER SPECIFICATIONS. HOWEVER, THE MAT WAS FOLDED TO FIT IN RETURN BOX BY THE COMPLAINANT. PER THE INSTRUCTIONS ON THE MAT, "DO NOT FOLD OR CREASE - STORE FLAT." SUBSEQUENT TO (B)(6) TESTING, (B)(6) WAS CONTACTED BY (B)(6) ABOUT THE ALLEGED INCIDENT ON (B)(6) 2011. THE RISK MANAGER AND DIRECTOR OF NURSING STATED THAT THEIR INTERNAL REPORT INDICATED THAT A MONITOR WAS IN USE WITH THE PATIENT AT THE TIME OF THE INCIDENT, BUT THAT THE MONITOR REPORTEDLY DID NOT ALARM DURING THE INCIDENT. HOWEVER, THE REPORT ALSO SAID THAT THE MONITOR DID ALARM JUST AFTER THE INCIDENT, WHEN THE RESIDENT WAS BEING RETURNED TO THE BED.

Description of Event or Problem · 1

ON (B)(6) 2011, (B)(6) REPORTED THAT HIS MOTHER HAD FALLEN IN (B)(6) 2010 WHILE RECEIVING CARE AT (B)(6). (B)(6) STATED THAT THE NURSE ON STAFF SAID THE MONITOR DID NOT ALARM WHEN THE FALL OCCURRED. (B)(6) STATED THAT THE BED X-RAY TAKEN BY (B)(6) HAD INDICATED THAT SHE WAS FINE. AFTER RETURNING HOME FOR A COUPLE OF DAYS, (B)(6) TOOK HIS MOTHER TO A DIFFERENT HOSPITAL WHERE THEY DISCOVERED THAT SHE HAD BROKEN HER PELVIS IN 2 PLACES. (B)(6) CONTACTED (B)(6) ON (B)(6) 2011 AND SHIPPED THE EQUIPMENT BACK ON (B)(6) 2011 FOR EVALUATION.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 MICRO-TECH INFORMER MATE AND PAD KMI STANLEY SECURITY SOLUTIONS, INC. 81781, 82710 30110 (82710)

Patients

Seq Age Sex Outcome Treatment
1 Hospitalization