5.0MM AGGRESSIVE PLUS CUTTER, FORMULA (5BX)
Report
- Report Number
- 2648666-2011-00145
- Event Type
- Malfunction
- Date Received
- May 27, 2011
- Date of Event
- April 29, 2011
- Report Date
- May 4, 2011
- Manufacturer
- STRYKER ENDOSCOPY PUERTO RICO
- Product Code
- HRX
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- AL, US
- Reporter Occupation
- NOT APPLICABLE
Narratives
THE REPORTED UNIT WAS NOT RETURNED FOR EVALUATION AND THE BROKEN TIP CONDITION COULD NOT BE CONFIRMED. NEVERTHELESS, TIP BREAKAGE IS A KNOWN FAILURE MODE, WHICH PROBABLE ROOT CAUSES, ASSOCIATED RISKS, AND MITIGATION MEASURES ARE CONTEMPLATED WITHIN THE RISK DOCUMENTATION. REVIEW OF THE DEVICE HISTORY RECORD (DHR) OF THE INVOLVED LOT AND PART NUMBER DISCLOSED NO DISCREPANCIES THAT COULD BE ASSOCIATED TO THE REPORTED FAILURE. BASED ON MENTIONED DOCUMENTATION AND EVALUATION OF UNITS RECEIVED FROM PREVIOUS SIMILAR EVENTS, MOST PROBABLE ROOT CAUSES FOR THIS CONDITION CAN BE ASSOCIATED, BUT NOT LIMITED TO: COMPONENTS DO NOT FIT PROPERLY (I.E ALIGNMENT / GAP BETWEEN CUTTER AND HOUSING ASSEMBLY). EXCESSIVE APPLIED PRESSURE. CONTACT OF THE SHAVER BLADE TIP AGAINST A METAL OBJECT. MATERIAL STRENGTH: EVEN THOUGH HISTORICAL DATA HAD SHOWN NO NON-CONFORMANCES ASSOCIATED TO THIS CONDITION, IT IS CURRENTLY BEING EVALUATED BASED ON FRACTURE MODE AND RECURRENCE OF SIMILAR EVENTS. SHOULD THE PRODUCT BE RETURNED, IT WILL BE EVALUATED AND A FOLLOW UP REPORT WILL BE FILED. THE FAILURE MODE WILL BE MONITORED FOR FUTURE REOCCURRENCE.
IT WAS REPORTED THAT THERE WAS A DELAY IN THE CASE DUE TO A BROKEN PIECE OF THE CUTTER UNIT. IT WAS FURTHER REPORTED THAT IT WAS REMOVED BY THE SURGEON. FURTHER, THE CASE WAS COMPLETED SUCCESSFULLY USING A DIFFERENT UNIT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | 5.0MM AGGRESSIVE PLUS CUTTER, FORMULA (5BX) | CUTTER | HRX | STRYKER ENDOSCOPY PUERTO RICO | 10335CE2 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | UNK |