Report
- Report Number
- 2023826-2011-00508
- Event Type
- Malfunction
- Date Received
- June 10, 2011
- Date of Event
- May 13, 2011
- Report Date
- May 17, 2011
- Manufacturer
- STAAR SURGICAL COMPANY
- Product Code
- HQL
- PMA / PMN Number
- P990013
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- PA, US
- Reporter Occupation
- OTHER
Narratives
COLLAMER ULTRAVIOLET-ABSORBING POSTERIOR CHAMBER SINGLE PIECE FOLDABLE INTRAOCULAR LENS. (B)(4). METHOD: LENS WORK ORDER SEARCH. RESULTS: VISUAL INSPECTION OF THE RETURNED PRODUCT FOUND ONE HAPTIC TORN. A PIECE OF THE LENS OPTIC AND THE OTHER HAPTIC WERE TORN OFF AND MISSING. THE LENS WAS RETURNED IN LIQUID AND THERE WAS EVIDENCE OF CLEAR SURGICAL RESIDUE. A LENS WORK ORDER SEARCH WAS PERFORMED AND NO SIMILAR COMPLAINTS WERE FOUND WITHIN THE WORK ORDER. CONCLUSIONS: BASED ON THE COMPLAINT HISTORY, WORK ORDER SEARCH AND THE EVALUATION OF THE RETURNED PRODUCT, A SPECIFIC ROOT CAUSE OF THE EVENT COULD NOT BE DETERMINED. (B)(4).
THE REPORTER STATED THE SURGEON INSERTED A (B)(4) COLLAMER ASPHERIC SINGLE PIECE LENS AND THE HAPTIC TORE. THE LENS WAS REMOVED WITH NO PATIENT INJURY, NO ENLARGED INCISION OR SUTURES. ANOTHER SAME MODEL LENS WAS IMPLANTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | SEE H-10 | INTRAOCULAR LENS | HQL | STAAR SURGICAL COMPANY | CC4204A |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 88 YR | INJECTOR SYSTEM MODEL NANOPOINT - LOT NUMBER UNK |