FDA Adverse Event Malfunction Summary report: N

SEE H-10

MDR report key: 2120357 · Received June 10, 2011

Report

Report Number
2023826-2011-00508
Event Type
Malfunction
Date Received
June 10, 2011
Date of Event
May 13, 2011
Report Date
May 17, 2011
Manufacturer
STAAR SURGICAL COMPANY
Product Code
HQL
PMA / PMN Number
P990013
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
PA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

COLLAMER ULTRAVIOLET-ABSORBING POSTERIOR CHAMBER SINGLE PIECE FOLDABLE INTRAOCULAR LENS. (B)(4). METHOD: LENS WORK ORDER SEARCH. RESULTS: VISUAL INSPECTION OF THE RETURNED PRODUCT FOUND ONE HAPTIC TORN. A PIECE OF THE LENS OPTIC AND THE OTHER HAPTIC WERE TORN OFF AND MISSING. THE LENS WAS RETURNED IN LIQUID AND THERE WAS EVIDENCE OF CLEAR SURGICAL RESIDUE. A LENS WORK ORDER SEARCH WAS PERFORMED AND NO SIMILAR COMPLAINTS WERE FOUND WITHIN THE WORK ORDER. CONCLUSIONS: BASED ON THE COMPLAINT HISTORY, WORK ORDER SEARCH AND THE EVALUATION OF THE RETURNED PRODUCT, A SPECIFIC ROOT CAUSE OF THE EVENT COULD NOT BE DETERMINED. (B)(4).

Description of Event or Problem · 1

THE REPORTER STATED THE SURGEON INSERTED A (B)(4) COLLAMER ASPHERIC SINGLE PIECE LENS AND THE HAPTIC TORE. THE LENS WAS REMOVED WITH NO PATIENT INJURY, NO ENLARGED INCISION OR SUTURES. ANOTHER SAME MODEL LENS WAS IMPLANTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 SEE H-10 INTRAOCULAR LENS HQL STAAR SURGICAL COMPANY CC4204A

Patients

Seq Age Sex Outcome Treatment
1 88 YR INJECTOR SYSTEM MODEL NANOPOINT - LOT NUMBER UNK