FDA Adverse Event Malfunction Summary report: N

BD Q-SYTE 15CM SMALL BORE BI-EXT SET

MDR report key: 21203262 · Received January 21, 2025

Report

Report Number
9610847-2025-00001
Event Type
Malfunction
Date Received
January 21, 2025
Date of Event
December 28, 2024
Report Date
February 5, 2025
Manufacturer
BD INFUSION THERAPY SYSTEMS INC. S.A. DE C.V.
Product Code
FPA
UDI-DI
00382903851577
PMA / PMN Number
K142527
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IN
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

H.3. IF A DEVICE EVALUATION AND/OR DEVICE HISTORY REVIEW IS COMPLETED, A SUPPLEMENTAL REPORT WILL BE FILED.

Additional Manufacturer Narrative · 0

INVESTIGATION RESULTS: A DEVICE HISTORY RECORD REVIEW WAS COMPLETED BY OUR QUALITY ENGINEER TEAM FOR PROVIDED MATERIAL NUMBER 385157 AND LOT NUMBER 4061404. THE REVIEW DID NOT REVEAL ANY DETECTED ABNORMALITIES DURING THE PRODUCTION PROCESS THAT COULD HAVE CONTRIBUTED TO THIS DEFECT AND ALL QUALITY TESTS WERE FOUND TO BE WITHIN SPECIFICATION. AS A SAMPLE WAS UNAVAILABLE FOR RETURN, A THOROUGH SAMPLE INVESTIGATION COULD NOT BE COMPLETED. BASED ON THE INVESTIGATION RESULTS, AN EXACT CAUSE FOR THIS INCIDENT COULD NOT BE IDENTIFIED. THERE ARE QUALITY CONTROLS CURRENTLY IN PLACE TO DETECT THIS TYPE OF DEFECT DURING THE PRODUCTION PROCESS. FURTHER ACTION HAS NOT BEEN DETERMINED NECESSARY AT THIS TIME. COMPLAINTS RECEIVED FOR THIS DEVICE AND REPORTED CONDITION WILL CONTINUE TO BE TRACKED AND TRENDED. OUR QUALITY TEAM REGULARLY REVIEWS THE COLLECTED DATA FOR IDENTIFICATION OF EMERGING TRENDS.

Description of Event or Problem · 0

IT WAS REPORTED THAT BD Q-SYTE 15CM SMALL BORE BI-EXT SET LEAKED DRUG LEAKAGE FROM Q-SYTE BI EXTENSION IN RADIOLOGY DEPARTMENT. THEY HAVE GIVEN CT CONTRAST AT 5.5ML PER SECOND WITH OVERALL VOLUME OF 80ML THROUGH PRESSURE INJECTOR. THEY HAVE NOTICED DRUG LEAKAGE FROM Q-SYTE.

Description of Event or Problem · 0

NO ADDITIONAL INFORMATION.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
682211 BD Q-SYTE 15CM SMALL BORE BI-EXT SET INTRAVASCULAR ADMINISTRATION SET FPA BD INFUSION THERAPY SYSTEMS INC. S.A. DE C.V. 4061404 00382903851577

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown