RESTORE PRIME ADVANCED
Report
- Report Number
- 3007566237-2011-04016
- Event Type
- Injury
- Date Received
- June 6, 2011
- Date of Event
- January 1, 2011
- Report Date
- May 6, 2011
- Manufacturer
- MEDTRONIC NEUROMODULATION
- Product Code
- LGW
- PMA / PMN Number
- P840001
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- UK
- Reporter Occupation
- PHYSICIAN
Narratives
(B)(4).
THE PT LOST STIMULATION 6 MONTHS AFTER IMPLANTABLE NEUROSTIMULATOR (INS) IMPLANT. IMPEDANCES WERE MEASURED AND ELECTRODE 1 WAS OUT OF RANGE. THE PT WAS SUCCESSFULLY REPROGRAMMED WITHOUT USING ELECTRODE 1. FIVE MONTHS LATER, THE PT LOST STIMULATION AGAIN AND ELECTRODES 0 AND 1 WERE BOTH OUT OF RANGE. THE PT WAS SUCCESSFULLY REPROGRAMMED USING ELECTRODES 2 AND 3. SEVEN MONTHS LATER, THE PT LOST STIMULATION AND ALL IMPEDANCES WERE >40,000 OHMS. DESPITE THIS, THE PT WAS SUCCESSFULLY REPROGRAMMED USING ELECTRODES 6 AND 7. THE INS WAS AT END OF LIFE AND WAS BEING REPLACED. AFTER REPLACEMENT, THE IMPEDANCES WERE MEASURED AND ALL WERE >40,000 OHMS AT 3V. THE LEAD CONFIGURATION AND EXTENSION WERE BOTH CHECKED AND CONFIRMED AS CORRECT. THE EXTENSION WAS REMOVED, WIPED WITH A DRY SWAB, AND REINSERTED, BUT THE IMPEDANCES WERE STILL OUT OF RANGE. THE LEAD WAS TESTED WITH A SNAP LID CABLE AND AN EXTERNAL NEUROSTIMULATOR AND THE IMPEDANCES WERE STILL OUT OF RANGE. THE VISIBLE PORTION OF THE EXTENSION WAS EXAMINED AND THERE WERE NO OBVIOUS CUTS OR BREAKS. THE LEAD WAS TESTED AGAIN AND THE PT WAS ABLE TO FEEL STIMULATION USING ELECTRODES 6 AND 7 WHEN THE INS WAS SET TO 1.7V EVEN THOUGH ALL IMPEDANCES WERE OUT OF RANGE. THE PT WAS UNABLE TO FEEL ANY STIMULATION USING ELECTRODES 0-3. A LEAD FRACTURE WAS SUSPECTED BUT NOT CONFIRMED. THE PT WAS NOT INJURED AND WAS RECEIVING THERAPEUTIC BENEFIT. ADDITIONAL INFO HAS BEEN REQUESTED, A FOLLOW-UP REPORT WILL BE SENT IF ADDITIONAL INFO BECOMES AVAILABLE. SEE ALSO MANUFACTURER'S REPORT # 3007566237201104017.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | RESTORE PRIME ADVANCED | LGW | MEDTRONIC NEUROMODULATION | 37702 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Required Intervention | EXPLANTED:| IMPLANTED:| LEAD: MODEL 3998, LOT# B0467370K| IMPLANTED:| EXTENSION: MODEL 37082, LOT# NKB007618N| EXPLANTED: |