THERAKOS UVAR XTS PHOTOPHERESIS SYSTEM
Report
- Report Number
- 2523595-2011-00017
- Event Type
- Injury
- Date Received
- June 3, 2011
- Date of Event
- May 24, 2011
- Report Date
- June 6, 2011
- Manufacturer
- THERAKOS
- Product Code
- LNR
- PMA / PMN Number
- P860003
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NY, US
- Reporter Occupation
- NURSE
Narratives
NO REPORTS OF A DEVICE MALFUNCTION HAVE BEEN RECEIVED TO DATE. A LOT FILE REVIEW WAS PERFORMED ON UVADEX LOT 1939641 (EXP: APR 2013). FORMULATION, FILLING, TERMINAL STERILIZATION, IN-PROCESS, AND FINAL QA TESTING ALL MET REQUIREMENTS. (B)(4).
ON (B)(6) 2011, THERAKOS RECEIVED A REPORT OF A PT WHO DEVELOPED A RASH AND SWELLING AFTER RECEIVING THEIR 5 ECP TREATMENT ON (B)(6) 2011 AND RETURNED TO THE HOSPITAL FOR OBSERVATION. THE PT SEZARY SYNDROME AND EXTENSIVE PATCH PLAQUE CTCL. PREVIOUSLY TREATED WITH MTX (NO EFFICACY) AND TARGRETIN (DIFFUSE RASH WITH MILD DESQUAMATION). THE PT WAS TREATED WITH ECP 3 WEEKS AGO AND FIRST 2 TREATMENTS WERE WITHOUT EVENT. AFTER SECOND 2 TREATMENTS, THE PT DEVELOPED DIFFUSE ERYTHEMA, HIVES, FACIAL SWELLING WITHOUT BREATHING DIFFICULTIES AND WAS TREATED WITH ORAL STEROIDS ANTIHISTAMINES WITH RESOLUTION OVER THE INTERVENING WK. THE PT RECEIVED ECP TREATMENT #5 AND DEVELOPED DIFFUSE ERYTHEMA, ITCHING, HIVES, FACIAL SWELLING WITHOUT BREATHING DIFFICULTIES AFTER GOING HOME. THE PT WAS SEEN IN THE ECP UNIT ON THE MORNING OF (B)(6) 2011 AND WAS NOT IN ANY RESPIRATORY DISTRESS ALTHOUGH UNCOMFORTABLE FROM SKIN FINDINGS. THE PT WILL BE STARTED ON STEROIDS AND IS BEING EVALUATED BY THE ALLERGIST. ALTHOUGH THE REPORTER CONSIDERS THE EVENT SERIOUS, THEY DO NOT BELIEVE THE EVENT IS NECESSARILY RELATED AS THE PT DEVELOPED HIVES AND ALL AGAIN OVER A WEEK LATER. THE PT IS NO LONGER RECEIVING ECP.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | THERAKOS UVAR XTS PHOTOPHERESIS SYSTEM | XTS ECP SYSTEM | LNR | THERAKOS |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 54 YR | Other | CLINDAMYCIN LOTION| BONIVA| KENALOG| HYDROXYZINE| ZANTAC| HYDROCODONE-HOMATROPINE| BENADRYL| CHLORAHEXADINE GLUCONATE| SARNA LOTION| CARMOL| ALBUTEROL| BENZONATATE| ELIDEL| NASONEX| CORDRAN| FLUOCINOLONE |