BD VACUTAINER® CPT¿ CELL PREPARATION TUBE WITH SODIUM CITRATE
Report
- Report Number
- 1917413-2024-01403
- Event Type
- Malfunction
- Date Received
- January 21, 2025
- Date of Event
- December 25, 2024
- Report Date
- December 31, 2024
- Manufacturer
- BECTON, DICKINSON & CO. (BROKEN BOW)
- Product Code
- JCF
- UDI-DI
- 30382903627609
- PMA / PMN Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CH
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
E1: INITIAL REPORTER FACILITY NAME: (B)(6). INVESTIGATION SUMMARY: BD RECEIVED SIX PHOTOS FOR INVESTIGATION. THE EVALUATION OF THESE PHOTOS REVEALED FLOATING REDDISH-BROWN STRANDS IN THE LIQUID DENSITY MEDIA (LDM) LAYER OF THE TUBE. THESE STRANDS ARE COMPOSED OF RESIDUAL SILICONE COATING AND TRACE METALS FROM THE LDM, WHICH ARE A RESULT OF THE MANUFACTURING PROCESS AND ARE NOT CONSIDERED FOREIGN MATTER AS THEY ARE INTRINSIC TO THE COMPONENTS OF THE TUBE. BASED ON A REVIEW OF THE DEVICE HISTORY RECORD FOR THE INCIDENT LOT, ALL PRODUCT SPECIFICATIONS AND REQUIREMENTS FOR LOT RELEASE WERE MET. THIS COMPLAINT HAS NOT BEEN CONFIRMED FOR THE INDICATED FAILURE MODE: FOREIGN MATTER - UNKNOWN. COMPLAINTS RECEIVED FOR THIS DEVICE AND REPORTED CONDITION WILL CONTINUE TO BE TRACKED AND TRENDED. INFORMATION WILL BE CAPTURED ON TREND REPORTS AND MONITORED. OUR BUSINESS TEAM REGULARLY REVIEWS THE COLLECTED DATA FOR THE IDENTIFICATION OF EMERGING TRENDS.
IT WAS REPORTED PRIOR TO USE OF BD VACUTAINER® CPT¿ CELL PREPARATION TUBE WITH SODIUM CITRATE 110 TUBES CONTAINED FOREIGN MATTER(FM). THERE WAS NO HEALTH IMPACT OR CONSEQUENCES REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 722487 | BD VACUTAINER® CPT¿ CELL PREPARATION TUBE WITH SODIUM CITRATE | LYMPHOCYTE SEPARATION MEDIUM | JCF | BECTON, DICKINSON & CO. (BROKEN BOW) | 4197918 | 30382903627609 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Unknown |