FDA Adverse Event Injury Summary report: N

RESTORE RECHARGEABLE NEUROSTIMULATOR

MDR report key: 2120021 · Received June 8, 2011

Report

Report Number
3004209178-2011-04135
Event Type
Injury
Date Received
June 8, 2011
Date of Event
May 10, 2011
Report Date
January 26, 2017
Manufacturer
MDT PUERTO RICO OPERATIONS CO., JUNCOS
Product Code
LGW
PMA / PMN Number
P840001
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MN, US
Reporter Occupation
NOT APPLICABLE
Health Professional
N

Narratives

Additional Manufacturer Narrative · 1

(B)(4). ANALYSIS RESULTS WERE NOT AVAILABLE AT THE TIME OF THIS REPORT. A F/U REPORT WILL BE SENT WHEN ANALYSIS IS COMPLETED.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE PT WAS EXPERIENCING BURNING IN THE POCKET SITE. IMPEDANCES WERE CHECKED WITH SEVERAL HIGH. LEAD REMOVED FROM THE INS AND CHECKED IN MLTC (MULTI LEAD TRIALING CABLE). ALL IMPEDANCES NORMAL IN MLTC. LEAD RECONNECTED TO INS. IMPEDANCES HIGH ON DIFFERENT ELECTRODES. LEADS DISCONNECTED AGAIN AND CHECKED IN MLTC WHERE ALL IMPEDANCES REMAINED NORMAL WITH MLTC BUT HIGH (AGAIN DIFFERENT ELECTRODES) WHEN CONNECTED TO INS. NEW INS OPENED AND LEADS CONNECTED WITH ALL IMPEDANCES NORMAL. THE PT WAS NO LONGER EXPERIENCING BURNING IN POCKET WITH THE NEW INS. THE PT RECOVERED WITHOUT SEQUELAE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 RESTORE RECHARGEABLE NEUROSTIMULATOR LGW MDT PUERTO RICO OPERATIONS CO., JUNCOS 37711 NA

Patients

Seq Age Sex Outcome Treatment
1 35 YR Female Required Intervention ACCESSORY: MODEL 37752, LOT# NKA145348N| EXPLANTED:| EXPLANTED:| EXPLANTED:| EXPLANTED:| EXPLANTED:| EXPLANTED:| EXPLANTED:| EXPLANTED:| EXTENSION: MODEL 37082, LOT# NKB005033V.| EXTENSION: MODEL 37082, LOT# NKB016108N.| IMPLANTED:| IMPLANTED:| IMPLANTED:| IMPLANTED:| IMPLANTED:| IMPLANTED:| IMPLANTED:| IMPLANTED:| LEAD: MODEL 3487A, LOT# V210506.| LEAD: MODEL 3487A, LOT# V537001.| LEAD: MODEL 3888 LOT# V536164.| LEAD: MODEL 3888, LOT# V536164.| N/A.| PROGRAMMER: MODEL 37743, LOT# NKE156037N.| LEAD: MODEL 3487A, LOT# V210506| EXPLANTED:| IMPLANTED:| IMPLANTED:| IMPLANTED:| IMPLANTED:| IMPLANTED:| PROGRAMMER: MODEL 37743, LOT# NKE156037N| EXPLANTED:| IMPLANTED:| EXTENSION: MODEL 37082, LOT# NKB016108N| EXPLANTED:| IMPLANTED:| EXPLANTED:| IMPLANTED:| EXPLANTED:| EXPLANTED:| EXPLANTED:| LEAD: MODEL 3888 LOT# V536164| LEAD: MODEL 3487A, LOT# V537001| EXTENSION: MODEL 37082, LOT# NKB005033V| LEAD: MODEL 3888, LOT# V536164| EXPLANTED:| ACCESSORY: MODEL 37752, LOT# NKA145348N