FDA Adverse Event
Injury
Summary report: N
RESTORE RECHARGEABLE NEUROSTIMULATOR
MDR report key: 2120021
·
Received June 8, 2011
Report
- Report Number
- 3004209178-2011-04135
- Event Type
- Injury
- Date Received
- June 8, 2011
- Date of Event
- May 10, 2011
- Report Date
- January 26, 2017
- Manufacturer
- MDT PUERTO RICO OPERATIONS CO., JUNCOS
- Product Code
- LGW
- PMA / PMN Number
- P840001
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MN, US
- Reporter Occupation
- NOT APPLICABLE
- Health Professional
- N
Narratives
Additional Manufacturer Narrative · 1
(B)(4). ANALYSIS RESULTS WERE NOT AVAILABLE AT THE TIME OF THIS REPORT. A F/U REPORT WILL BE SENT WHEN ANALYSIS IS COMPLETED.
Description of Event or Problem · 1
IT WAS REPORTED THAT THE PT WAS EXPERIENCING BURNING IN THE POCKET SITE. IMPEDANCES WERE CHECKED WITH SEVERAL HIGH. LEAD REMOVED FROM THE INS AND CHECKED IN MLTC (MULTI LEAD TRIALING CABLE). ALL IMPEDANCES NORMAL IN MLTC. LEAD RECONNECTED TO INS. IMPEDANCES HIGH ON DIFFERENT ELECTRODES. LEADS DISCONNECTED AGAIN AND CHECKED IN MLTC WHERE ALL IMPEDANCES REMAINED NORMAL WITH MLTC BUT HIGH (AGAIN DIFFERENT ELECTRODES) WHEN CONNECTED TO INS. NEW INS OPENED AND LEADS CONNECTED WITH ALL IMPEDANCES NORMAL. THE PT WAS NO LONGER EXPERIENCING BURNING IN POCKET WITH THE NEW INS. THE PT RECOVERED WITHOUT SEQUELAE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | RESTORE RECHARGEABLE NEUROSTIMULATOR | LGW | MDT PUERTO RICO OPERATIONS CO., JUNCOS | 37711 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 35 YR | Female | Required Intervention | ACCESSORY: MODEL 37752, LOT# NKA145348N| EXPLANTED:| EXPLANTED:| EXPLANTED:| EXPLANTED:| EXPLANTED:| EXPLANTED:| EXPLANTED:| EXPLANTED:| EXTENSION: MODEL 37082, LOT# NKB005033V.| EXTENSION: MODEL 37082, LOT# NKB016108N.| IMPLANTED:| IMPLANTED:| IMPLANTED:| IMPLANTED:| IMPLANTED:| IMPLANTED:| IMPLANTED:| IMPLANTED:| LEAD: MODEL 3487A, LOT# V210506.| LEAD: MODEL 3487A, LOT# V537001.| LEAD: MODEL 3888 LOT# V536164.| LEAD: MODEL 3888, LOT# V536164.| N/A.| PROGRAMMER: MODEL 37743, LOT# NKE156037N.| LEAD: MODEL 3487A, LOT# V210506| EXPLANTED:| IMPLANTED:| IMPLANTED:| IMPLANTED:| IMPLANTED:| IMPLANTED:| PROGRAMMER: MODEL 37743, LOT# NKE156037N| EXPLANTED:| IMPLANTED:| EXTENSION: MODEL 37082, LOT# NKB016108N| EXPLANTED:| IMPLANTED:| EXPLANTED:| IMPLANTED:| EXPLANTED:| EXPLANTED:| EXPLANTED:| LEAD: MODEL 3888 LOT# V536164| LEAD: MODEL 3487A, LOT# V537001| EXTENSION: MODEL 37082, LOT# NKB005033V| LEAD: MODEL 3888, LOT# V536164| EXPLANTED:| ACCESSORY: MODEL 37752, LOT# NKA145348N |