FDA Adverse Event Malfunction Summary report: N

TEMP PUMP TEMPERATURE

MDR report key: 2119661 · Received May 27, 2011

Report

Report Number
1313850-2011-00119
Event Type
Malfunction
Date Received
May 27, 2011
Date of Event
May 2, 2011
Report Date
May 2, 2011
Manufacturer
STRYKER CORP DBA GAYMAR
Product Code
FOH
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
ID, US
Reporter Occupation
OTHER

Narratives

Description of Event or Problem · 1

IT WAS REPORTED THE UNIT WAS NOT HEATING TO TEMPERATURE. UPON EVAL, THERE WAS EVIDENCE OF FLUID INGRESS TO THE PC BOARD. NO PT INVOLVEMENT OR ADVERSE CONSEQUENCES WERE REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 TEMP PUMP TEMPERATURE PUMP FOH STRYKER CORP DBA GAYMAR TP500 NA

Patients

Seq Age Sex Outcome Treatment
1