ALINITY I STAT HIGH SENSITIVE TROPONIN-I REAGENT KIT
Report
- Report Number
- 3005094123-2025-00022
- Event Type
- Malfunction
- Date Received
- January 17, 2025
- Date of Event
- December 10, 2024
- Report Date
- February 28, 2025
- Manufacturer
- ABBOTT IRELAND DIAGNOSTICS DIVISION
- Product Code
- MMI
- PMA / PMN Number
- N/A
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- AS
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
THIS REPORT IS BEING FILED ON AN INTERNATIONAL PRODUCT, LIST NUMBER 8P13 THAT HAS A SIMILAR PRODUCT DISTRIBUTED IN THE US, LIST NUMBER 4Z21 AND 510K/PMA/BLA K202525. SECTION A. PATIENT INFORMATION OF MULTIPLE IS FOR 2 PATIENTS. SID (B)(6) IS 51 YEAR OLD MALE, NO GENDER, RACE, ETHNICITY PROVIDED. SID (B)(6) IS 78 YEAR OLD MALE, NO GENDER, RACE, ETHNICITY PROVIDED. AN EVALUATION IS IN PROCESS. A FOLLOW-UP REPORT WILL BE SUBMITTED WHEN THE EVALUATION IS COMPLETE.
SECTION D4. CATALOG WAS UPDATED FROM 8P13-22 ON INITIAL SUBMISSION TO 8P13-37 ON FOLLOW-UP SUBMISSION. SECTION D4 PRIMARY UDI NUMBER WAS UPDATED TO (B)(4). THIS REPORT IS BEING FILED ON AN INTERNATIONAL PRODUCT, LIST NUMBER 8P13 THAT HAS A SIMILAR PRODUCT DISTRIBUTED IN THE US, LIST NUMBER 4Z21 AND 510K/PMA/BLA K202525. AN EVALUATION IS IN PROCESS. A FOLLOW-UP REPORT WILL BE SUBMITTED WHEN THE EVALUATION IS COMPLETE.
SECTION D4 EXPIRATION DATE WAS UPDATED TO 2/5/2025. PRIMARY UDI NUMBER WAS UPDATED TO (B)(4). THE COMPLAINT INVESTIGATION INCLUDED A REVIEW OF DATA AND INFORMATION PROVIDED BY THE CUSTOMER, SEARCH FOR SIMILAR COMPLAINTS, TICKET TRENDING REVIEW, LABELING REVIEW, DEVICE HISTORY RECORD REVIEW AND FIELD DATA REVIEW. RETURN TESTING WAS NOT COMPLETED AS RETURNS WERE NOT AVAILABLE. DATA AND INFORMATION PROVIDED BY THE CUSTOMER WERE REVIEWED AND SUPPORT THE COMPLAINT ISSUE WITHOUT INDICATION FOR ANY ADDITIONAL ISSUE. REVIEW OF TRACKING AND TRENDING DATA FOR ALINITY I STAT HIGH SENSITIVE TROPONIN-I ASSAY DID NOT IDENTIFY AN INCREASE IN COMPLAINTS RELATED TO THE COMPLAINT ISSUE. ADDITIONALLY, THE SIMILAR COMPLAINT TICKET SEARCH FOR LOT 67109UD00 DID NOT IDENTIFY AN INCREASE IN COMPLAINT ACTIVITY. LABELING WAS REVIEWED AND SUFFICIENTLY ADDRESSES THE CUSTOMER¿S ISSUE. DEVICE HISTORY REVIEW DID NOT IDENTIFY ANY NON-CONFORMANCES OR DEVIATIONS WITH THE COMPLAINT LOT AND COMPLAINT ISSUE. CUSTOMER FIELD DATA WAS USED TO ASSESS THE PERFORMANCE OF ALINITY I STAT HIGH SENSITIVE TROPONIN-I ASSAY AND LOT 67109UD00 USING WORLDWIDE DATA. THE REVIEW OF APPLICABLE FIELD DATA SHOWS PERFORMANCE DATA IS ACCEPTABLE. BASED ON THE INVESTIGATION, NO SYSTEMIC ISSUE OR DEFICIENCY OF THE ALINITY I STAT HIGH SENSITIVE TROPONIN-I REAGENT LOT 67109UD00 WAS IDENTIFIED.
THE ACCOUNT GENERATED FALSE ELEVATED ALINITY I STAT HIGH SENSITIVE TROPONIN-I ON 2 PATIENT SAMPLES THAT REPEATED LOWER. ON 10DEC 2024, SID (B)(6) (51 YEAR OLD MALE) GENERATED ALINITY I STAT HIGH SENSITIVE TROPONIN-I OF 746 NG/L (NEAT) BUT REPEATED LOWER <2 NG/L (HARD SPIN). ON (B)(6) 2025; SID (B)(6); (78 YEAR OLD MALE) GENERATED ALINITY I STAT HIGH SENSITIVE TROPONIN-I OF 141.2 NG/L (NEAT) BUT REPEATED LOWER 3.8, 4.5, 4.1 NG/L (HARD SPIN). NO CUSTOMER TROPONIN CUTOFF WAS PROVIDED. NO GENDER, RACE, ETHNICITY WAS PATIENT INFORMATION WAS PROVIDED. NO IMPACT TO PATIENT MANAGEMENT WAS PROVIDED.
THE ACCOUNT GENERATED FALSE ELEVATED ALINITY I STAT HIGH SENSITIVE TROPONIN-I ON 2 PATIENT SAMPLES THAT REPEATED LOWER. ON (B)(6) 2024, SID (B)(6) (51 YEAR OLD MALE) GENERATED ALINITY I STAT HIGH SENSITIVE TROPONIN-I OF 746 NG/L (NEAT) BUT REPEATED LOWER <2 NG/L (HARD SPIN). ON (B)(6) 2025 SID (B)(6) (78 YEAR OLD MALE) GENERATED ALINITY I STAT HIGH SENSITIVE TROPONIN-I OF 141.2 NG/L (NEAT) BUT REPEATED LOWER 3.8, 4.5, 4.1 NG/L (HARD SPIN). NO CUSTOMER TROPONIN CUTOFF WAS PROVIDED. NO GENDER, RACE, ETHNICITY WAS PATIENT INFORMATION WAS PROVIDED. NO IMPACT TO PATIENT MANAGEMENT WAS PROVIDED.
THE ACCOUNT GENERATED FALSE ELEVATED ALINITY I STAT HIGHLY SENSITIVE TROPONIN-I ON 2 PATIENT SAMPLES THAT REPEATED LOWER. ON (B)(6) 2024, SID (B)(6) (51 YEAR OLD MALE) GENERATED ALINITY I STAT HIGHLY SENSITIVE TROPONIN-I OF 746 NG/L (NEAT) BUT REPEATED LOWER <2 NG/L (HARD SPIN). ON (B)(6) 2025 SID (B)(6) (78 YEAR OLD MALE) GENERATED ALINITY I STAT HIGHLY SENSITIVE TROPONIN-I OF 141.2 NG/L (NEAT) BUT REPEATED LOWER 3.8, 4.5, 4.1 NG/L (HARD SPIN). NO CUSTOMER TROPONIN CUTOFF WAS PROVIDED. NO GENDER, RACE, ETHNICITY WAS PATIENT INFORMATION WAS PROVIDED. NO IMPACT TO PATIENT MANAGEMENT WAS PROVIDED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1144955 | ALINITY I STAT HIGH SENSITIVE TROPONIN-I REAGENT KIT | IMMUNOASSAY METHOD, TROPONIN SUBUNIT | MMI | ABBOTT IRELAND DIAGNOSTICS DIVISION | 67109UD00 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Unknown | ALNTY I PROCESSING MODU, 03R65-01, (B)(6).| ALNTY I PROCESSING MODU, 03R65-01, (B)(6).| ALNTY I PROCESSING MODU, 03R65-01, (B)(6). |