FDA Adverse Event Death Summary report: N

BENEVISION DISTRIBUTED MONITORING SYSTEM (DMS)

MDR report key: 21183840 · Received January 17, 2025

Report

Report Number
3009156722-2025-00002
Event Type
Death
Date Received
January 17, 2025
Date of Event
December 20, 2024
Report Date
December 3, 2025
Manufacturer
SHENZHEN MINDRAY BIO-MEDICAL ELECTRONICS CO., LTD
Product Code
MSX
PMA / PMN Number
K220058
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IN, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

SYSTEM LOGS AND DATA ASSOCIATED WITH THE OCCURRENCE WERE EVALUATED, AND NO MALFUNCTION WAS CONFIRMED FOR ANY DEVICES INVOLVED IN THE REPORT. THE REPORTED WIDENED QRS COMPLEXES ARE THE REGULAR VENTRICULAR-PACED RHYTHM AT THE TIME OF THE OCCURRENCE, WHICH DIDN'T MEET THE ALARM THRESHOLDS. THE ARRHYTHMIA ALARM FUNCTION IS PERFORMED PER SPECIFICATION. CORRECTION: B3 (CORRECTING THE YEAR DATE).

Additional Manufacturer Narrative · 0

SYSTEM LOGS AND DATA ASSOCIATED WITH THE OCCURRENCE WERE EVALUATED, AND NO MALFUNCTION WAS CONFIRMED FOR ANY DEVICES INVOLVED IN THE REPORT. THE REPORTED WIDENED QRS COMPLEXES ARE THE REGULAR VENTRICULAR-PACED RHYTHM AT THE TIME OF THE OCCURRENCE, WHICH DIDN'T MEET THE ALARM THRESHOLDS. THE ARRHYTHMIA ALARM FUNCTION IS PERFORMED PER SPECIFICATION.

Description of Event or Problem · 0

IT WAS REPORTED THAT ON DECEMBER (B)(6) 2024, APPROXIMATELY BETWEEN 11:00 AM AND 12:30 PM, IN ROOM ID A510, ON A PATIENT PLACED ON THE TM80 TELEMETRY MONITOR (ID TELE50) THE CENTRAL MONITORING UNIT (CMU) DEPARTMENT MONITOR TECHNICIAN NOTED THE PATIENT WAS IN PACED RHYTHM AT RATE 70 AT 11:00 AM. THE CMU WAS NOTIFIED BY GATEWAY STAFF THAT THE PATIENT WAS BEING RESUSCITATED. THE MONITOR TECHNICIAN NOTED THAT QRS HAD WIDENED AND THAT THERE WAS NO ALARM. THE PATIENT EXPIRED.

Description of Event or Problem · 0

IT WAS REPORTED THAT ON (B)(6) 2024, APPROXIMATELY BETWEEN 11:00 AM AND 12:30 PM, IN ROOM ID (B)(6), ON A PATIENT PLACED ON THE TM80 TELEMETRY MONITOR (ID TELE50) THE CENTRAL MONITORING UNIT (CMU) DEPARTMENT MONITOR TECHNICIAN NOTED THE PATIENT WAS IN PACED RHYTHM AT RATE 70 AT 11:00 AM. THE CMU WAS NOTIFIED BY GATEWAY STAFF THAT THE PATIENT WAS BEING RESUSCITATED. THE MONITOR TECHNICIAN NOTED THAT QRS HAD WIDENED AND THAT THERE WAS NO ALARM. THE PATIENT EXPIRED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1465495 BENEVISION DISTRIBUTED MONITORING SYSTEM (DMS) BENEVISION WORK STATION MSX SHENZHEN MINDRAY BIO-MEDICAL ELECTRONICS CO., LTD BENEVISION WORK STATION

Patients

Seq Age Sex Outcome Treatment
1 NA Female Death