FDA Adverse Event Injury Summary report: N

SCARLET ALT SECURED LUMBAR ANTERIOR CAGE - HYPERLORDOTIC

MDR report key: 21178699 · Received January 17, 2025

Report

Report Number
3007728266-2025-00001
Event Type
Injury
Date Received
January 17, 2025
Date of Event
January 1, 2024
Report Date
January 17, 2025
Manufacturer
SPINEART SA
Product Code
OVD
UDI-DI
07640185342071
Report Source
Manufacturer report
Reporter Location
SZ
Reporter Occupation
501

Narratives

Additional Manufacturer Narrative · 0

AFTER RECEPTION OF THE COMPLAINT, THE MANUFACTURING FOLDER HAS BEEN REVIEWED. IT IS CONFIRMED THAT THE RAW MATERIALS AND PRODUCTION PROCESSES ARE CONFORM TO THE PREDEFINED SPECIFICATIONS. THERE HAVE BEEN NO NON-CONFORMITIES OBSERVED DURING THE MANUFACTURING PROCESS. COMPLAINED DEVICE BELONGS TO A BATCH OF (B)(4) UNITS, MANUFACTURED IN (B)(6) 2021. THIS IS THE 1ST COMPLAINT WE RECEIVED FOR THIS BATCH. SCIENTIFIC LITERATURE REPORTS THAT VENTRAL HERNIA (SIMILAR TO EVENTRATION) IS A COMMON COMPLICATION OF ABDOMINAL SURGERIES (SEE "NHO" AND "CARO" ARTICLES - CPT FOLDER). BASED ON THE ABOVE INFORMATION, WE CONCLUDE THAT THE ISSUE REPORTED IS NOT LINKED TO THE DEVICE IMPLANTED BUT THAT IT IS RELATED TO THE PROCEDURE. SINCE THE LAUNCH OF THE DEVICE, (B)(4) CAGES (SCARLET ALT AND SCARLET ALT HL RANGES) HAVE BEEN IMPLANTED AND THIS IS THE 8TH COMPLAINT WE RECEIVE FOR THIS KIND OF ISSUE, (B)(4), THE RATE IS VERY LOW. WE CLOSE THE CASE AS IS WITH NO FURTHER ACTION.

Description of Event or Problem · 0

ON (B)(6) 2025, WE RECEIVED A COMPLAINT FROM THE FIELD ((B)(6)), FROM CLINICAL STUDIES, (SEE (B)(4) FOLDER) REPORTING THAT, AROUND 7 MONTHS POSTOPERATIVE, THE PATIENT SUFFERED FROM AN ABDOMINAL WALL EVENTRATION. THE SITE CLASSIFIED THIS EVENT AS POSSIBLY RELATED TO THE PROCEDURE. THE PATIENT UNDERWENT A PARIETAL REPAIR SURGERY ON (B)(6) 2024. THE PATIENT IS DOING FINE. WE REPORT THIS CASE BECAUSE A DEVICE MARKETED IN THE US IN INVOLVED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1146531 SCARLET ALT SECURED LUMBAR ANTERIOR CAGE - HYPERLORDOTIC SCARLET ALT SECURED LUMBAR ANTERIOR CAGE - HYPERLORDOTIC OVD SPINEART SA SCA-LM 20 16-S 6-3418 07640185342071

Patients

Seq Age Sex Outcome Treatment
1 58 YR Female