PARALLAX CLEARVIEW ACCESS NEEDLE
Report
- Report Number
- 2951580-2011-00075
- Event Type
- Other
- Date Received
- June 1, 2011
- Date of Event
- September 2, 2010
- Report Date
- June 1, 2011
- Manufacturer
- ARTHROCARE CORP.
- Product Code
- KDD
- PMA / PMN Number
- EXEMPT
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CO, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
SINCE THE DEVICE WAS NOT RETURNED FOR INVESTIGATION, A COMPLETE INVESTIGATION COULD NOT BE PERFORMED. A LOT HISTORY REVIEW WAS PERFORMED. NO ABNORMALITIES WERE FOUND IN THE LOT HISTORY REVIEW. THE DEVICE WAS USED FOR DELIVERY OF STEM CELLS AND PLATELET RICH PLASMA TO THE PT'S HIP; HOWEVER, THE DEVICE IS INDICATED FOR DELIVERY OF BONE CEMENT FOR FIXATION OF FRACTURES OF THE VERTEBRAL BODY IN VERTEBROPLASTY AND KYPHOPLASTY PROCEDURES. THE CANNULA HAS ONLY BEEN DESIGNED FOR THE INDICATED USE; ANY OTHER USE HAS NOT BEEN EVALUATED BY ARTHROCARE. THE IFU PROVIDES THE FOLLOWING INDICATION FOR USE: "THE PARALLAX CLEAR-VIEW ACCESS NEEDLES ARE INTENDED FOR USE WITH THE CEMENT DELIVERY SYSTEMS FOR THE FIXATION OF PATHOLOGICAL FRACTURES OF THE VERTEBRAL BODY USING VERTEBROPLASTY OR KYPHOPLASTY PROCEDURES. PAINFUL VERTEBRAL COMPRESSION FRACTURES MAY RESULT FROM OSTEOPOROSIS, BENIGN LESIONS (HEMANGIOMA), AND MALIGNANT LESIONS (METASTATIC CANCER, MYELOMA)".
DURING AN AVASCULAR NECROSIS (AVN) HIP PROCEDURE, THE PARALLAX CLEARVIEW ACCESS NEEDLE BROKE OFF IN THE PT'S HIP. THE NEEDLE WAS USED TO DELIVER STEM CELLS AND PLATELET RICH PLASMA (PRP) UNDER FLUOROSCOPY TO THE PT'S HIP. THE PROCEDURE WAS PERFORMED IN AN OUT-PT FACILITY AND WERE NOT EQUIPPED TO REMOVE THE NEEDLE FROM THE PT'S HIP AND THE PT WAS REFERRED TO AN IN-PT FACILITY FOR THE REMOVAL OF THE BROKEN NEEDLE WITH AN ORTHOPEDIC SURGEON. THE REMOVAL OF THE NEEDLE WAS PERFORMED THE FOLLOWING DAY.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | PARALLAX CLEARVIEW ACCESS NEEDLE | CEMENT DISPENSER | KDD | ARTHROCARE CORP. | M200753 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 74 YR | Other |