FDA Adverse Event Malfunction Summary report: N

COBAS E 801 IMMUNOASSAY ANALYZER

MDR report key: 21168132 · Received January 16, 2025

Report

Report Number
1823260-2025-00163
Event Type
Malfunction
Date Received
January 16, 2025
Date of Event
December 19, 2024
Report Date
February 7, 2025
Manufacturer
ROCHE DIAGNOSTICS
Product Code
JJE
UDI-DI
04015630946198
PMA / PMN Number
K162606
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MD, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

THE PTH REAGENT LOT NUMBER WAS 801714 WITH AN EXPIRATION DATE OF 30-SEP-2025.

Additional Manufacturer Narrative · 0

QC WAS ACCEPTABLE. CALIBRATION DATA FROM 21-DEC-2024 WAS UNACCEPTABLE. NO CALIBRATION DATA WAS PROVIDED FROM BEFORE THE EVENT. NO ISSUES WERE IDENTIFIED DURING A REVIEW OF THE ALARM TRACE DATA. BEFORE THE SERVICE VISIT, THE CUSTOMER REMOVED THE INCORRECT REAGENT BOTTLE AND LOADED A NEW PRECLEAN SYSTEM REAGENT. THE FIELD SERVICE ENGINEER (FSE) PERFORMED SIPPER AIR PURGES, SYSTEM AIR PURGES, SYSTEM PRIMES, REAGENT PRIMES, AND MEASURING CELL CONDITIONING. THE FSE OBSERVED THE INSTRUMENT OPERATING NORMALLY. THE CUSTOMER HAD NO FURTHER ISSUES. THE INVESTIGATION DETERMINED THE EVENT WAS DUE TO AN INCORRECT MAINTENANCE PROCEDURE PERFORMED BY THE CUSTOMER.

Description of Event or Problem · 0

THE CUSTOMER STATED A TECHNICIAN PLACED A DILUENT REAGENT IN THE PLACE OF THE PRECLEAN SYSTEM REAGENT ON A COBAS E 801 IMMUNOASSAY ANALYZER. THE CUSTOMER THINKS THIS MAY HAVE BEEN DONE ON THE MORNING OF (B)(6) 2024, HOWEVER, THIS COULD NOT BE CONFIRMED. THE CUSTOMER REPEATED 473 PATIENT SAMPLES ON A DIFFERENT E801 ANALYZER. THE RESULTS FOR 3 PATIENTS WERE CORRECTED. THE RESULTS FOR 1 PATIENT TESTED FOR ELECSYS PTH (PTH) WERE DISCREPANT. THE INITIAL PTH RESULT WAS 60.6 PG/ML. THE REPEAT RESULT WAS 41.8 PG/ML. THE REPEAT RESULT WAS BELIEVED TO BE CORRECT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1792014 COBAS E 801 IMMUNOASSAY ANALYZER CLINICAL IMMUNOASSAY ANALYZER JJE ROCHE DIAGNOSTICS E801 04015630946198

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown