VENTED AUTOFEED CHAMBER
Report
- Report Number
- 9611451-2011-00334
- Event Type
- Malfunction
- Date Received
- June 7, 2011
- Date of Event
- May 11, 2011
- Report Date
- May 11, 2011
- Manufacturer
- FISHER & PAYKEL HEALTHCARE LTD
- Product Code
- BTT
- PMA / PMN Number
- K934140
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MN, US
- Reporter Occupation
- OTHER
Narratives
(B)(4). METHOD: TWO COMPLAINT CHAMBERS WERE RETURNED, WITH LOT NUMBERS 100307 AND 101127 RESPECTIVELY. THE RETURNED COMPLAINT CHAMBERS WERE ATTACHED TO A WATERBAG FOR TESTING. RESULTS: ONE OF THE CHAMBERS FILLED TO THE EXPECTED FILL LEVEL AND THEN SMALL DROPS OF WATER BEGAN TO BUILD UP AT THE CONNECTION BETWEEN THE FEEDSET TUBE AND THE WATER BAG SPIKE. WITH THE SECOND CHAMBER NO LEAKS WERE FOUND EVEN WHEN THE WATER FEEDSET WAS MANIPULATED. A LOT CHECK REVEALED NO OTHER COMPLAINTS FOR EITHER LOT NUMBER. CONCLUSION: THE LEAK WAS CAUSED BY THE FEEDSET TUBING NOT BEING GLUED PROPERLY AT THE SPIKE, DUE TO INSUFFICIENT GLUE APPLIED BY THE GLUING MACHINE. ALL CHAMBERS ARE PRESSURE TESTED BEFORE THEY LEAVE THE PRODUCTION LINE AND ANY HOLES OR LEAKS IN THE FEEDSET ARE IDENTIFIED DURING THIS PROCESS. THIS SUGGESTS THAT THE LEAK DEVELOPED POST PRODUCTION. THE USER INSTRUCTIONS WHICH ACCOMPANY THE MR290 CHAMBER STATE "PERFORM A PRESSURE AND LEAK TEST ON THE BREATHING SYSTEM AND CHECK FOR OCCLUSIONS BEFORE CONNECTING TO A PATIENT". AS PART OF OUR ONGOING PRODUCT IMPROVEMENTS, ADDITIONAL GLUE IS NOW APPLIED TO THE FEEDSET SPIKE DURING PRODUCTION. (B)(4).
A HOSPITAL IN (B)(6) REPORTED THAT TWO MR290 AUTOFEED HUMIDIFICATION CHAMBERS WERE LEAKING AT THE WATER BAG SPIKE. NO PATIENT CONSEQUENCE WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | VENTED AUTOFEED CHAMBER | AUTOFEED CHAMBER | BTT | FISHER & PAYKEL HEALTHCARE LTD | MR290V | 100307 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |