FDA Adverse Event Injury Summary report: N

HYPE SCL 2 LATERALIZED CEMENTLESS STEM

MDR report key: 21156449 · Received January 15, 2025

Report

Report Number
3008668801-2025-00014
Event Type
Injury
Date Received
January 15, 2025
Date of Event
December 17, 2024
Report Date
February 4, 2026
Manufacturer
S.E.R.F SOCIETE DETUDES RECHERCHE FABRICATION
Product Code
LZO
PMA / PMN Number
K223745
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
FR
Reporter Occupation
OTHER
Health Professional
N

Narratives

Additional Manufacturer Narrative · 0

REPORTED EVENT : AN EVENT REGARDING SUBSIDENCE (LEADING TO REVISION) INVOLVING AN HYPE SCL 2 LATERALIZED CEMENTLESS STEM WAS REPORTED. CONCLUSIONS : THE REPORTED DEVICE IS AN SERF PRODUCT. NO NON-COMPLIANCE OBSERVED IN THIS BATCH 20 UNITS MARKETED BY SERF IN OCTOBER 2011 A COMPLAINT WAS RECORDED ON THIS BATCH RC-24-0408 AS PART OF THE SAME CLINICAL STUDY BUT FOR A DIFFERENT EVENT WHOSE CAUSE COULD NOT BE ESTABLISHED. LACK OF TECHNICAL INVESTIGATION: NO PRODUCT RETURN - COMPLAINT RELATED TO A CLINICAL STUDY IN THE ABSENCE OF FURTHER INFORMATION, THE CAUSE CANNOT BE ESTABLISHED. CORRECTIVE ACTION/PREVENTIVE ACTION: NO ACTION IS REQUIRED BY STRYKER AT THIS TIME AS THE INVESTIGATION FOR THIS EVENT IS PERFORMED BY SERF.

Additional Manufacturer Narrative · 0

AN EVALUATION OF THE DEVICE CANNOT BE PERFORMED AS THE DEVICE WAS NOT RETURNED TO THE MANUFACTURER. SHOULD ADDITIONAL INFORMATION BECOME AVAILABLE IT WILL BE REPORTED IN A SUPPLEMENTAL REPORT UPON COMPLETION OF THE INVESTIGATION.

Additional Manufacturer Narrative · 0

MANUFACTURER ENTITY D3 CORRECTED TO S.E.R.F SOCIETE DETUDES RECHERCHE FABRICATION.

Description of Event or Problem · 0

SERF REPORTED RECEIVING RESULTS OF A CLINICAL STUDY. SUBSIDENCE AND ROTATION OF THE FEMORAL STEM (PAIN).

Description of Event or Problem · 0

SERF REPORTED RECEIVING RESULTS OF A CLINICAL STUDY. SUBSIDENCE AND ROTATION OF THE FEMORAL STEM (PAIN).

Description of Event or Problem · 0

SERF REPORTED RECEIVING RESULTS OF A CLINICAL STUDY. SUBSIDENCE AND ROTATION OF THE FEMORAL STEM (PAIN).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1274280 HYPE SCL 2 LATERALIZED CEMENTLESS STEM PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED LZO S.E.R.F SOCIETE DETUDES RECHERCHE FABRICATION 1107148A

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown Hospitalization| R