AXIOS STENT AND ELECTROCAUTERY ENHANCED DELIVERY SYSTEM
Report
- Report Number
- 3005099803-2024-06920
- Event Type
- Injury
- Date Received
- January 14, 2025
- Date of Event
- December 29, 2022
- Report Date
- January 14, 2025
- Manufacturer
- BOSTON SCIENTIFIC CORPORATION
- Product Code
- PCU
- PMA / PMN Number
- K220112
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TW
- Reporter Occupation
- PHYSICIAN
- Health Professional
- Yes
Narratives
INITIAL REPORTER ADDRESS NO. 21, SECTION 2, NANYA S. ROAD; REPORTED HERE AS THE ADDRESS EXCEEDED CHARACTER LIMIT FOR THE DESIGNATED FIELD. LITERATURE SOURCE: CHEN-SHUAN CHUNG ET AL. RESCUING MALDEPLOYMENT OF LUMEN-APPOSING METAL STENT FOR THE MANAGEMENT OF ACUTE CHOLECYSTITIS. DIGESTIVE ENDOSCOPY (2023) 35:670. DOI: 10.1111/DEN.14572. IMDRF DEVICE CODE A1502 CAPTURES THE REPORTABLE EVENT OF STENT FIRST FLANGE DIFFICULT TO POSITION.
BOSTON SCIENTIFIC CORPORATION BECAME AWARE OF THE FOLLOWING EVENT FROM REFERENCED LITERATURE ARTICLE "RESCUING MALDEVELOPMENT OF LUMEN-APPOSING METAL STENT FOR THE MANAGEMENT OF ACUTE CHOLECYSTITIS." ACCORDING TO THE LITERATURE, AN AXIOS STENT AND ELECTROCAUTERY ENHANCED DELIVERY SYSTEM WAS IMPLANTED TRANSDUODENAL TO GALLBLADDER FOR THE TREATMENT OF ACUTE CHOLECYSTITIS DURING AN ENDOSCOPIC ULTRASOUND-GUIDED GALLBLADDER DRAINAGE PROCEDURE PERFORMED ON (B)(6) 2022. DURING THE PROCEDURE, THE STENT WAS DEPLOYED; HOWEVER, NO DRAINAGE OF BILE WAS OBSERVED. IT WAS CONFIRMED UNDER FLUOROSCOPY THAT THE DISTAL FLANGE WAS DEPLOYED BETWEEN THE DUODENUM AND GALLBLADDER. A GUIDEWIRE WAS THEN ADVANCED THROUGH THE LUMEN-APPOSING METAL STENT AND THREE 10F-3CM AND TWO 7F-3CM DOUBLE PIGTAIL PLASTIC STENTS WERE SUCCESSFULLY IMPLANTED. NO PATIENT COMPLICATIONS WERE REPORTED DUE TO THIS EVENT AND THE PATIENT LIVED UNEVENTFULLY FOR 6 MONTHS. NOTE: IT WAS REPORTED THAT THE AXIOS STENT WAS PLACED TRANSDUODENAL TO GALLBLADDER. HOWEVER, PER THE AXIOS STENT AND ELECTROCAUTERY- ENHANCED DELIVERY SYSTEM DIRECTIONS FOR USE, THE STENT IS INDICATED FOR USE TO FACILITATE TRANSGASTRIC OR TRANSDUODENAL ENDOSCOPIC DRAINAGE OF A PANCREATIC PSEUDOCYST OR A WALLED-OFF NECROSIS WITH >= 70 % FLUID CONTENT. THE DEVICE IS NOT INDICATED TO BE IMPLANTED IN THE GALLBLADDER.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1516734 | AXIOS STENT AND ELECTROCAUTERY ENHANCED DELIVERY SYSTEM | PANCREATIC STENT, COVERED, METALLIC, REMOVABLE | PCU | BOSTON SCIENTIFIC CORPORATION | M00553570 | 0027516261 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 85 YR | Female | Required Intervention |