FDA Adverse Event Malfunction Summary report: N

CERENOVUS ENTERPRISE

MDR report key: 21150206 · Received January 14, 2025

Report

Report Number
3008114965-2025-00059
Event Type
Malfunction
Date Received
January 14, 2025
Date of Event
December 30, 2024
Report Date
February 4, 2025
Manufacturer
MEDOS INTERNATIONAL SARL
Product Code
NJE
UDI-DI
10886704075349
PMA / PMN Number
H60001
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CH
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

PRODUCT COMPLAINT # (B)(4). INFORMATION REGARDING PATIENT WEIGHT, HEIGHT, MEDICAL HISTORY, RACE, AND ETHNICITY WAS NOT REPORTED. SECTION E1. INITIAL REPORTER PHONE: (B)(6). SECTION H3 - THE DEVICE IS AVAILABLE TO BE RETURNED FOR EVALUATION AND TESTING. HOWEVER, IT HAS NOT BEEN RECEIVED TO DATE. IF THE DEVICE RETURNS, A DEVICE INVESTIGATION WILL BE PERFORMED. MISSING INFORMATION FROM THIS REPORT IS IDENTIFIED AS BLANK; THIS INFORMATION WAS NOT PROVIDED IN THE REPORTED EVENT OR AVAILABLE AT THE TIME OF REPORT SUBMISSION. THIS REPORT IS BEING SUBMITTED PURSUANT TO THE PROVISIONS OF 21 CFR, PART 803 (AND/OR PART 4, AS APPLICABLE). THIS REPORT MAY BE BASED ON INFORMATION WHICH HAS NOT BEEN INVESTIGATED OR VERIFIED PRIOR TO THE REQUIRED REPORTING DATE. THIS REPORT DOES NOT REFLECT A CONCLUSION BY CERENOVUS, OR ITS EMPLOYEES THAT THE REPORT CONSTITUTES AN ADMISSION THAT THE PRODUCT, CERENOVUS, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE POTENTIAL EVENT DESCRIBED IN THIS REPORT. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL REPORT, A FOLLOW-UP REPORT WILL BE FILED AS APPROPRIATE.

Additional Manufacturer Narrative · 0

PRODUCT COMPLAINT # (B)(4). UPDATED SECTIONS ON THIS MEDWATCH: B4, D9, G3, G6, H2, H3, H6 AND H11. COMPLAINT CONCLUSION: AS REPORTED BY THE FIELD, DURING AN ENDOVASCULAR EMBOLIZATION, AN ENTERPRISE2 4MMX16MM NO TIP INTRACRANIAL STENT (ENCR401600, 9042461) BECAME IMPEDED IN PROXIMAL OF AN UNSPECIFIED MICROCATHETER (MC) AND COULD NOT ADVANCE ANY MORE. THE PHYSICIAN TRIED TO RETRACT THE STENT ALONE, THE STENT WAS FOUND DETACHED FROM THE DELIVERY WIRE IN THE MICROCATHETER. THE PHYSICIAN USED A MICRO-GUIDE WIRE TO REMOVE THE STENT OUT AND SWITCH TO A NEW STENT TO COMPLETE THE SURGERY. THE MICROCATHETER WAS NOT REPLACED. THERE WAS NO PATIENT INJURY REPORTED. ADDITIONAL EVENT INFORMATION RECEIVED ON (B)(6) 2025 INDICTED THAT THEY WERE ABLE TO TORQUE THE DEVICE. THERE WAS NO EVIDENCE OF PHYSICAL MATERIAL WITHIN THE DEVICE. THE MICROCATHETER DID NOT KINK/BENT. THERE WAS NO EXCESSIVE FORCE USED WITH THE DEVICES. THE PROCEDURE WAS DELAYS BY 15 MINUTES DUE TO THE EVENT, THE DELAY WAS NOT CLINICALLY SIGNIFICANT. A NON-STERILE ENTERPRISE2 4MMX16MM NO TIP WAS RECEIVED CONTAINED IN THE DECONTAMINATION POUCH. UPON RECEIVING THE DEVICE, VISUAL INSPECTION WAS PERFORMED, AND THE STENT WAS DETACHED FROM THE DELIVERY UNIT. MICROSCOPIC INSPECTION WAS PERFORMED ON THE STENT COMPONENT. IT WAS OBSERVED TO BE IN GOOD CONDITION; THERE WAS NO STRUCTURAL DAMAGE (I.E., NO BROKEN STRUTS, NO KINKS); ALSO, IT WAS NOTED FULLY EXPANDED, BOTH ENDS CAN BE NOTED AS COMPLETELY FLARED. THE DELIVERY WIRE WAS SUBJECTED TO DIMENSIONAL ANALYSIS AND ALL MEASUREMENTS WERE FOUND TO BE WITHIN SPECIFICATION, INCLUDING THOSE SPECIFICATIONS THAT CONTROL THE ATTACHMENT AND DELIVERY OF THE STENT. THEREFORE, DEVICE FAILURE IS NOT SUSPECTED TO BE A CONTRIBUTING FACTOR. THE CUSTOMER COMPLAINT REGARDING A STENT BEING DETACHED WAS CONFIRMED SINCE THE STENT WAS NOTED AS ALREADY SEPARATED FROM THE DELIVERY SYSTEM; BASED ON THIS CONDITION, THE ISSUE REGARDING A STENT BEING IMPEDED IN PROXIMAL OF THE CONCOMITANT MICROCATHETER CANNOT BE EVALUATED THROUGH FUNCTIONAL TESTING. THE STENT MUST BE INSIDE THE INTRODUCER TUBE TO PERFORM THE FUNCTIONAL ANALYSIS. ADDITIONALLY, NONE OF THE RETURNED COMPONENTS PRESENT DAMAGES THAT SUGGEST THAT THEY WERE FORCIBLY ADVANCED. WITH THE LIMITED INFORMATION AVAILABLE, A CONCLUSIVE CAUSE CANNOT BE DETERMINED; HOWEVER, IT IS POSSIBLE THAT CLINICAL AND PROCEDURAL FACTORS, INCLUDING DEVICE MANIPULATION AND OPERATOR'S TECHNIQUE, MAY HAVE CONTRIBUTED TO THE REPORTED FAILURE. AT THIS TIME, THERE IS NO EVIDENCE TO SUPPORT THAT THE ISSUE REPORTED IN THE COMPLAINT IS A RESULT OF A DEFECT INHERENTLY RELATED TO THE DEVICE. LAKE REGION MEDICAL DID REVIEW THE DEVICE HISTORY RECORDS RELATIVE TO THE MANUFACTURING, INSPECTING AND PACKAGING OF THE LOT 9042461. THE HISTORY RECORDS INDICATE THIS PRODUCT WAS FINAL INSPECTION TESTED AT LAKE REGION MEDICAL AND WAS DETERMINED TO BE ACCEPTABLE. AS PART OF THE J&J MEDTECH QUALITY PROCESS, ALL DEVICES ARE MANUFACTURED, INSPECTED, AND RELEASED TO APPROVED SPECIFICATIONS. SINCE THERE WAS NO EVIDENCE TO SUGGEST THE EVENT WAS RELATED TO A MANUFACTURING OR DESIGN ISSUE, NO CAPA ACTIVITY IS REQUIRED. IT SHOULD BE NOTED THAT PRODUCT FAILURE COULD BE CAUSED BY MULTIPLE FACTORS. THE INSTRUCTIONS FOR USE (IFU) DO CONTAIN THE FOLLOWING RECOMMENDATIONS: ¿ DO NOT PARTIALLY DEPLOY THE STENT FROM THE INTRODUCER. ¿ CONFIRM THAT THE DELIVERY WIRE DOES NOT MOVE RELATIVE TO THE INTRODUCER DURING THE REMOVAL OF THE ENTERPRISE VASCULAR RECONSTRUCTION DEVICE AND DELIVERY SYSTEM FROM THE DISPENSER HOOP. ¿ CONFIRM THE TIP OF THE DELIVERY WIRE IS ENTIRELY WITHIN THE INTRODUCER. AS PART OF THE POST MARKET SURVEILLANCE PROGRAM, INFORMATION FROM THIS COMPLAINT IS TRENDED TO IDENTIFY STATISTICAL SIGNALS FOR CONSIDERATION OF FURTHER CORRECTION ACTION. SINCE THERE WAS NO EVIDENCE TO SUGGEST THE EVENT WAS RELATED TO A MANUFACTURING OR DESIGN ISSUE, NO CORRECTIVE ACTIONS WILL BE TAKEN AT THIS TIME. A SUPPLEMENTAL REPORT WILL BE SUBMITTED IF NEW FACTS ARISE WHICH MATERIALLY ALTER INFORMATION SUBMITTED IN A PREVIOUS MDR REPORT.

Description of Event or Problem · 0

AS REPORTED BY THE FIELD, DURING AN ENDOVASCULAR EMBOLIZATION, AN ENTERPRISE2 4MMX16MM NO TIP INTRACRANIAL STENT (ENCR401600, 9042461) BECAME IMPEDED IN PROXIMAL OF AN UNSPECIFIED MICROCATHETER (MC) AND COULD NOT ADVANCE ANY MORE. THE DOCTOR TRIED TO RETRACT THE STENT ALONE, THE STENT WAS FOUND DETACHED FROM THE DELIVERY WIRE IN THE MICROCATHETER. THE DOCTOR USED A MICRO-GUIDE WIRE TO REMOVE THE STENT OUT AND SWITCH TO A NEW STENT TO COMPLETE THE SURGERY. THE MICROCATHETER WAS NOT REPLACED. THERE WAS NO PATIENT INJURY REPORTED. ADDITIONAL EVENT INFORMATION RECEIVED ON 06-JAN-2025 INDICTED THAT THEY WERE ABLE TO TORQUE THE DEVICE. THERE WAS NO EVIDENCE OF PHYSICAL MATERIAL WITHIN THE DEVICE. THE MICROCATHETER DID NOT KINK/BENT. THERE WAS NO EXCESSIVE FORCE USED WITH THE DEVICES. THE PROCEDURE WAS DELAYS BY 15 MINUTES DUE TO THE EVENT, THE DELAY WAS NOT CLINICALLY SIGNIFICANT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1485935 CERENOVUS ENTERPRISE INTRACRANIAL NEUROVASCULAR STENT NJE MEDOS INTERNATIONAL SARL 9042461 10886704075349

Patients

Seq Age Sex Outcome Treatment
1 NA Female MICROCATHETER (UNSPECIFIED BRAND).