FDA Adverse Event Injury Summary report: N

CARPENTIER-EDWARDS PERIMOUNT THEON PERICARDIAL BIOPROSTHESIS

MDR report key: 21149805 · Received January 14, 2025

Report

Report Number
2015691-2025-00321
Event Type
Injury
Date Received
January 14, 2025
Date of Event
December 10, 2024
Report Date
February 13, 2025
Manufacturer
EDWARDS LIFESCIENCES
Product Code
DYE
UDI-DI
00690103194043
PMA / PMN Number
P860057
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
WI, US
Reporter Occupation
PHYSICIAN
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

H11: ADDITIONAL MANUFACTURER NARRATIVE: UPDATED SECTIONS G3, H6 (TYPE OF INVESTIGATION, INVESTIGATION FINDINGS, INVESTIGATION CONCLUSIONS). REGURGITATION IDENTIFIED POST-PROCEDURALLY HAS MANY POTENTIAL ROOT CAUSES, INCLUDING STRUCTURAL VALVE DETERIORATION (SVD), NON-STRUCTURAL VALVE DYSFUNCTION (NSVD), ENDOCARDITIS AND/OR THROMBOSIS. REGURGITATION OCCURRING DURING DEVICE IMPLANTATION OR POST-PROCEDURALLY IS MOST COMMONLY RELATED TO PATIENT AND/OR PROCEDURAL FACTORS AND IS NOT TYPICALLY AN INDICATION OF A DEVICE MALFUNCTION RELATED TO A MANUFACTURING NON-CONFORMANCE. THERE IS NO EVIDENCE TO SUGGEST A MANUFACTURING DEFECT CAUSED OR CONTRIBUTED TO THE REPORTED EVENT. THE MOST LIKELY ROOT CAUSE IS PATIENT-RELATED FACTORS, INCLUDING HYPERLIPIDEMIA (HLD), DYSLIPIDEMIA. THE INSTRUCTIONS FOR USE (IFU) HAVE BEEN REVIEWED, AND NO INADEQUACIES HAVE BEEN IDENTIFIED WITH REGARDS TO WARNINGS, CONTRAINDICATIONS, AND THE DIRECTIONS/CONDITIONS FOR THE SUCCESSFUL USE OF THE DEVICE. THE REPORTED TYPE OF EVENT IS INCLUDED IN THE IFU. THE DEVICE HISTORY RECORD (DHR) REVIEW WAS COMPLETED, AND THIS DEVICE PASSED ALL MANUFACTURING AND STERILIZATION INSPECTIONS PRIOR TO RELEASE FOR DISTRIBUTION. THERE WERE NO ISSUES IDENTIFIED THAT WOULD HAVE IMPACTED THIS EVENT. ALL PERTINENT INFORMATION AVAILABLE TO EDWARDS LIFESCIENCES HAS BEEN SUBMITTED. EDWARDS WILL CONTINUE TO REVIEW AND MONITOR ALL REPORTED EVENTS. TRENDS ARE MONITORED ON A MONTHLY BASIS AND IF ACTION IS REQUIRED, APPROPRIATE INVESTIGATION WILL BE PERFORMED. H11: CORRECTED DATA: CORRECTED SECTION H6 (COMPONENT CODE).

Additional Manufacturer Narrative · 0

H11: ADDITIONAL MANUFACTURER NARRATIVE: THE SUBJECT DEVICE IS NOT AVAILABLE FOR EVALUATION AS IT REMAINS IMPLANTED IN THE PATIENT. THE INVESTIGATION IS STILL IN PROGRESS; THEREFORE, A CONCLUSION HAS YET TO BE ESTABLISHED. A SUPPLEMENTAL REPORT WILL BE SUBMITTED ACCORDINGLY UPON INVESTIGATION COMPLETION. EDWARDS WILL CONTINUE TO REVIEW AND MONITOR ALL REPORTED EVENTS. TRENDS ARE MONITORED ON A MONTHLY BASIS AND IF ACTION IS REQUIRED, APPROPRIATE INVESTIGATION WILL BE PERFORMED.

Description of Event or Problem · 0

IT WAS LEARNED THROUGH IMPLANT PATIENT REGISTRY THAT A PATIENT WITH A 25MM 2700TFX AORTIC VALVE UNDERWENT A VALVE-IN-VALVE PROCEDURE AFTER AN IMPLANT DURATION OF SEVEN (7) YEARS, EIGHT (8) MONTHS DUE TO SEVERE REGURGITATION. THE PATIENT PRESENTED WITH SYMPTOMS OF HEART FAILURE NYHA CLASS II. THE PROCEDURE WAS PERFORMED WITH A 26MM 9755RSL TRANSCATHETER VALVE. EDWARDS LIFESCIENCES MAINTAINS AN IMPLANT PATIENT REGISTRY. THIS REGISTRY IS A PATIENT TRACKING MECHANISM FOR SERIALIZED EDWARDS IMPLANTABLE DEVICES (BIOPROSTHETIC HEART VALVES AND ANNULOPLASTY RINGS), RATHER THAN A TRUE POST-MARKET SURVEILLANCE REGISTRY. THROUGH THE REGISTRY, EDWARDS IS NOTIFIED WHEN THESE DEVICES ARE IMPLANTED. IN ADDITION, PATIENT AND/OR DEVICE STATUS MAY BE REPORTED TO THE REGISTRY VIA THE IMPLANTATION DATA CARDS. THE INFORMATION IS RECEIVED FROM VARIOUS SOURCES (E.G. SURGEON, HOSPITAL, AND PATIENT FAMILY MEMBERS) AND IS NOT RECEIVED IN THE FORM OF A CONVENTIONAL 'CUSTOMER COMPLAINT'. THE INFORMATION REPORTED MAY OR MAY NOT BE RELATED TO THE EDWARDS DEVICE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1898156 CARPENTIER-EDWARDS PERIMOUNT THEON PERICARDIAL BIOPROSTHESIS REPLACEMENT HEART VALVE DYE EDWARDS LIFESCIENCES 2700TFX 00690103194043

Patients

Seq Age Sex Outcome Treatment
1 81 YR Male Required Intervention| H| L