LIPASE
Report
- Report Number
- 2050012-2011-02081
- Event Type
- Malfunction
- Date Received
- June 6, 2011
- Date of Event
- May 13, 2011
- Report Date
- May 13, 2011
- Manufacturer
- BECKMAN COULTER BIOMEDICAL LIMITED
- Product Code
- CDT
- PMA / PMN Number
- K963424
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- GM
- Reporter Occupation
- OTHER
Narratives
PER CUSTOMER, THIS ISSUE RE-OCCURS EVERY FEW WEEKS, WHEREBY THE LABORATORY WILL OBTAIN A SAMPLE RESULT WHICH DISPLAYS ATYPICAL LIPASE VALUES. THE SAMPLING ORDER AND CONTAMINATION PARAMETERS HAVE BEEN CHECKED BY THE CUSTOMER CONTACT AND ARE CORRECT. THE SPECIMEN WAS RETURNED TO BCI (CO. CLARE) FOR INVESTIGATION. TESTING WAS CONDUCTED OVER THREE LOTS OF LIPASE REAGENT, INCLUDING THE REAGENT LOTS USED BY THE CUSTOMER AT THE TIME OF THIS EVENT. THE RESULTS FROM BOTH, THE AU681-03E AND THE REFERENCE INSTRUMENT ACROSS THE THREE REAGENT LOTS ARE PROVIDED BELOW. A ROOT CAUSE FOR THIS EVENT HAS NOT BEEN DETERMINED. ANALYZER: AU681-03E, RG LOT 9830: 20.57, RG LOT 1026: 17.76, RG LOT 1173: 16.18; ALTERNATE INSTRUMENT, 21.15, 18.83, 16.65. UNITS: U/L.
A CUSTOMER CONTACTED BECKMAN COULTER INC., (BCI) IN REGARDS TO AN ERRONEOUSLY HIGH LIPASE RESULT OF 171.0U/L GENERATED BY THE AU681-03E CHEMISTRY ANALYZER. THE SAMPLE WAS REPEATED TWO MORE TIMES ON THIS INSTRUMENT AND THE FOLLOWING RESULTS WERE OBTAINED: 14.1U/L AND 218.2U/L. THE ELEVATED RESULT WAS NOT REPORTED OUT OF THE LAB. THERE WAS NO EFFECT TO PATIENT TREATMENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | LIPASE | LIPASE REAGENT | CDT | BECKMAN COULTER BIOMEDICAL LIMITED | NA | 1026 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |