FDA Adverse Event Injury Summary report: N

INSPIRA SMOOTH SILICONE GEL FILLED BREAST IMPLANT

MDR report key: 21146143 · Received January 13, 2025

Report

Report Number
9617229-2025-00730
Event Type
Injury
Date Received
January 13, 2025
Date of Event
December 16, 2024
Report Date
May 30, 2025
Manufacturer
ALLERGAN (COSTA RICA)
Product Code
FTR
PMA / PMN Number
P020056
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
LA, US
Reporter Occupation
PHYSICIAN
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

DEVICE EVALUATION: THE DEVICE RELATED TO THE REPORTED EVENT OF CAPSULAR CONTRACTURE AND MALPOSITION WAS RECEIVED ON APRIL 23, 2025 WITH LOT NUMBER 1203346. BASED ON THE DEVICE ANALYSIS GRID, THE ASSESSMENTS OF THE COMPLAINT ARE: CAPSULAR CONTRACTURE: UNABLE TO OBSERVE AS IT IS NOT RELATED TO THE MANUFACTURING PROCESS. MALPOSITION: UNABLE TO OBSERVE AS IT IS NOT RELATED TO THE MANUFACTURING PROCESS. AS PER THE INVESTIGATION PROCEDURE, CREASES WERE COMPLETED AND NONE OF THE OBSERVATIONS ARE FOUND TO BE POTENTIALLY RELATED TO THE MANUFACTURING PROCESS, NO FURTHER ACTIONS ARE REQUIRED.

Additional Manufacturer Narrative · 0

A REVIEW OF THE DEVICE HISTORY RECORD HAS BEEN COMPLETED. NO DEVIATIONS OR NON-CONFORMANCES NOTED. THE EVENT OF "CAPSULAR CONTRACTURE" IS A PHYSIOLOGICAL COMPLICATION AND ANALYSIS OF THE DEVICE GENERALLY DOES NOT ASSIST ALLERGAN IN DETERMINING A PROBABLE CAUSE FOR THIS EVENT. FURTHER INFORMATION FROM THE REPORTER REGARDING EVENT, PRODUCT, OR PATIENT DETAILS HAS BEEN REQUESTED. NO ADDITIONAL INFORMATION IS AVAILABLE AT THIS TIME. REASON FOR REOPERATION: CAPSULAR CONTRACTURE, BAKER GRADE III.

Additional Manufacturer Narrative · 0

ADDITIONAL, CHANGED, AND/OR CORRECTED DATA: B5, D6B, H6.

Description of Event or Problem · 0

HEALTHCARE PROFESSIONAL REPORTED LEFT SIDE CAPSULAR CONTRACTURE BAKER GRADE III AND "IMPLANTS DROPPING AND IMPLANTS REMAIN VERY HIGH". DEVICE HAS BEEN EXPLANTED AND REPLACED.

Description of Event or Problem · 0

HEALTHCARE PROFESSIONAL REPORTED LEFT SIDE CAPSULAR CONTRACTURE BAKER GRADE III AND "IMPLANTS DROPPING AND IMPLANTS REMAIN VERY HIGH". DEVICE REMAINS IMPLANTED.

Description of Event or Problem · 0

DEVICE HAS BEEN EXPLANTED AND REPLACED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1255426 INSPIRA SMOOTH SILICONE GEL FILLED BREAST IMPLANT PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED FTR ALLERGAN (COSTA RICA) 1203346

Patients

Seq Age Sex Outcome Treatment
1 42 YR Female Required Intervention