FDA Adverse Event Malfunction Summary report: N

BD INTIMA-II Y 24GAX0.75IN PRN/PRN SLM NPVC

MDR report key: 21139351 · Received January 13, 2025

Report

Report Number
3002601200-2024-00816
Event Type
Malfunction
Date Received
January 13, 2025
Date of Event
December 20, 2024
Report Date
February 18, 2025
Manufacturer
BD SUZHOU (MDS)
Product Code
FOZ
UDI-DI
00382903830800
PMA / PMN Number
UNK
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CH
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

H.3. IF A DEVICE EVALUATION AND/OR DEVICE HISTORY REVIEW IS COMPLETED, A SUPPLEMENTAL REPORT WILL BE FILED.

Additional Manufacturer Narrative · 0

1.DHR/BHR REVIEW LOT#3080107 1-THIS BATCH OF PRODUCTS WERE ASSEMBLED AT INTIMA II AUTO LINE 4 IN APRIL, 2023, AND PACKAGED AT R245 PACKAGE LINE IN APRIL, 2023. BATCH SIZE IS 19000 EA. 2-REVIEW THE IN-PROCESS TEST REPORTS AND OUTGOING TEST REPORTS, AND ALL TEST RESULTS MEET THE PRODUCT SPECIFICATIONS. 3-REVIEW THE PRODUCTION RECORDS WITH NO NONCONFORMANCE, DEVIATION OR REWORK ACTIVITIES. 4-THE PRN BATCHES USED IN THIS BATCH OF PRODUCTS ARE 308307 & 3083081, REVIEW THE RAW MATERIAL INSPECTION RECORDS, NO ABNORMALITIES. 5) CHECKED ALL COMPLAINT LOT OF THE BATCH RECORD, RAW MATERIAL PROCESSES NO CHANGES. THE STERILIZATION PROCESS IS NORMAL, THE BI STERILITY TEST PASSED, AND THE EO RESIDUE TEST PASSED, THE PRODUCT MET THE REQUIREMENT OF BI STERILITY TEST BEFORE RELEASED, SEE THE ATTACHMENT 383080-3080107 COC 2.THE CUSTOMER RETURNED 4 SAMPLES. THE SAMPLE GAUGE IS 24G, P/N IS 383080, BATCH NUMBER IS 3080107. ALL THE 4PCS SAMPLES HAVE NOT BEEN OPENED. THE SLEEVE STOPPERS AT THE TOP OF PRNS ARE BROKEN AND WHITENED (NON-FOREIGN MATTER). SEE ATTACHMENT PR# (B)(4) FOR THE PHOTO. 3.CHECK THE RETAINED SAMPLES OF THIS BATCH, AND NO COMPLAINT DEFECTS ARE FOUND, PLEASE SEE THE ATTACHED PR# (B)(4). 4. WHITENED ISSUE AND DAMAGE OF LATEX PLUG OF PRN IS A KIND OF AGING PHENOMENON.OBSERVE THE RETAINED PRODUCTS AND STORE THEM UNDER CONSTANT TEMPERATURE AND HUMIDITY, NO AGING PHENOMENON HAS BEEN FOUND IN 3 YEARS.ACCORDING TO PREVIOUS EXPERIENCE, HIGH TEMPERATURE, HIGH LIGHT INTENSITY, STRONG OXIDATION OR HUMID ENVIRONMENT CAN CAUSE PRODUCT OXIDATION FAILURE. THE ENVIRONMENT IS RELATIVELY STABLE IN THE PRODUCTION PROCESS, WHICH WILL NOT CAUSE THE AGING PHENOMENON. THE DEFECT MAY BE RELATED TO TRANSPORT AND STORAGE CONDITIONS. 5. NO SIMILAR COMPLAINTS HAVE BEEN RECEIVED FROM OTHER HOSPITALS ABOUT THIS BATCH OF PRODUCTS. CONCLUSION NO ABNORMALITIES ARE FOUND IN THE PROCESS AND RETAINED SAMPLES, AND NO SIMILAR COMPLAINTS HAVE BEEN RECEIVED FROM OTHER HOSPITALS ABOUT THIS BATCH OF PRODUCTS. HIGH TEMPERATURE, HIGH LIGHT INTENSITY, STRONG OXIDATION OR HUMID ENVIRONMENT CAN CAUSE PRODUCT OXIDATION FAILURE. THE ENVIRONMENT IS RELATIVELY STABLE IN THE PRODUCTION PROCESS, WHICH WILL NOT CAUSE THE AGING PHENOMENON, SO THE DEFECT MAY BE RELATED TO TRANSPORT AND STORAGE CONDITIONS. PLANT WILL CONTINUE TRACKING THIS KIND OF COMPLAINT DEFECT.

Description of Event or Problem · 0

IT WAS REPORTED THAT BD INTIMA-II Y 24GAX0.75IN PRN/PRN SLM NPVC HAD MOLD. BEFORE UNPACKING, I FOUND THAT THE HEPARIN CAP WAS MOLDY. I NEED TO FILE A CLAIM, A COMPLAINT RESPONSE LETTER, AND A COMPLAINT ACCEPTANCE LETTER. I CAN RETURN 10 DEFECTIVE PRODUCTS, AND I HAVE PHOTOS TO PROVIDE.

Description of Event or Problem · 0

NO ADDITIONAL INFORMATION PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
452281 BD INTIMA-II Y 24GAX0.75IN PRN/PRN SLM NPVC INTRAVASCULAR CATHETER FOZ BD SUZHOU (MDS) 3080107 00382903830800

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown