FDA Adverse Event Malfunction Summary report: N

CERENOVUS ENTERPRISE

MDR report key: 21138549 · Received January 13, 2025

Report

Report Number
3008114965-2024-01272
Event Type
Malfunction
Date Received
January 13, 2025
Date of Event
December 26, 2024
Report Date
March 5, 2025
Manufacturer
MEDOS INTERNATIONAL SARL
Product Code
NJE
UDI-DI
10878528008894
PMA / PMN Number
H60001
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CH
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

MANUFACTURER¿S REF. NO: (B)(4). THE PURPOSE OF THIS MDR SUBMISSION IS TO REPORT THAT MULTIPLE ATTEMPTS TO OBTAIN ADDITIONAL INFORMATION RELATED TO THE PROCEDURE, THE REPORTED DEVICE ISSUE, AND TO OBTAIN THE PRODUCT FOR ANALYSIS WERE UNSUCCESSFUL. IF ADDITIONAL INFORMATION IS AVAILABLE AT A LATER DATE AND THE PRODUCT IS RETURNED, THE FILE WILL BE UPDATED ACCORDINGLY. LAKE REGION MEDICAL PERFORMED A REVIEW OF THE DEVICE HISTORY RECORDS RELATIVE TO THE MANUFACTURING, INSPECTION, AND PACKAGING OF THE LOT: 9061786. THE HISTORY RECORD INDICATES THIS PRODUCT WAS FINAL INSPECTION TESTED AT LAKE REGION MEDICAL AND WAS DETERMINED TO BE ACCEPTABLE. THE PRODUCT INVESTIGATION WILL BE CLOSED AS NO FURTHER INVESTIGATION CAN BE PERFORMED AT THIS TIME. IF THE PRODUCT IS RETURNED OR ADDITIONAL INFORMATION IS PROVIDED AT A LATER DATE, THE FILE WILL BE UPDATED AND A SUPPLEMENTAL 3500A REPORT WILL BE SUBMITTED. WITH THE LIMITED INFORMATION AVAILABLE AND WITHOUT THE PRODUCT AVAILABLE FOR ANALYSIS, THE REPORTED ISSUE DOCUMENTED IN THE COMPLAINT COULD NOT BE CONFIRMED. BASED ON THE MANUFACTURING DOCUMENTATION REVIEW, THERE IS NO INDICATION THAT THE EVENT IS RELATED TO THE DEVICE MANUFACTURING PROCESS. THE EXACT CAUSE OF THE EVENT COULD NOT BE CONCLUSIVELY DETERMINED; HOWEVER, IT IS POSSIBLE THAT CIRCUMSTANCES OF THE PROCEDURE AND / OR DEVICE MANIPULATION / INTERACTION MAY HAVE CONTRIBUTED TO THE REPORTED FAILURE. AS PART OF THE POST MARKET SURVEILLANCE PROGRAM, INFORMATION FROM THIS COMPLAINT IS TRENDED FOR STATISTICAL SIGNALS AND CORRECTIVE / PREVENTIVE ACTION MAY BE TRIGGERED LATER. SINCE THERE WAS NO EVIDENCE TO SUGGEST THE EVENT WAS RELATED TO A MANUFACTURING OR DESIGN ISSUE, NO CORRECTIVE ACTIONS WILL BE TAKEN AT THIS TIME. THE MANUFACTURER WILL SUBMIT A SUPPLEMENTAL REPORT IF NEW FACTS ARISE WHICH MATERIALLY ALTER INFORMATION SUBMITTED IN A PREVIOUS MDR REPORT. ADDITIONAL INFORMATION WILL BE SUBMITTED WITHIN 30 DAYS OF RECEIPT.

Additional Manufacturer Narrative · 0

MANUFACTURER¿S REF. NO: (B)(4). INFORMATION REGARDING PATIENT IDENTIFIER, DATE OF BIRTH, AGE, SEX, GENDER, WEIGHT, RACE, AND ETHNICITY WERE NOT PROVIDED. SECTION E.1: THE INITIAL REPORTER PHONE: (B)(6). SECTION H.3: THE DEVICE IS AVAILABLE TO BE RETURNED FOR EVALUATION AND TESTING. HOWEVER, IT HAS NOT BEEN RECEIVED TO DATE. IF THE DEVICE RETURNS, A DEVICE INVESTIGATION WILL BE PERFORMED. LAKE REGION MEDICAL PERFORMED A REVIEW OF THE DEVICE HISTORY RECORDS RELATIVE TO THE MANUFACTURING, INSPECTION, AND PACKAGING OF THE LOT 9061786. THE HISTORY RECORD INDICATES THIS PRODUCT WAS FINAL INSPECTION TESTED AT LAKE REGION MEDICAL AND WAS DETERMINED TO BE ACCEPTABLE. THIS REPORT IS BEING SUBMITTED PURSUANT TO THE PROVISIONS OF 21 CFR, PART 803. THIS REPORT MAY BE BASED ON INFORMATION WHICH HAS NOT BEEN INVESTIGATED OR VERIFIED PRIOR TO THE REQUIRED REPORTING DATE. THIS REPORT DOES NOT REFLECT A CONCLUSION BY JOHNSON & JOHNSON MEDTECH, OR ITS EMPLOYEES THAT THE REPORT CONSTITUTES AN ADMISSION THAT THE PRODUCT, JOHNSON & JOHNSON MEDTECH, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE POTENTIAL EVENT DESCRIBED IN THIS REPORT. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL REPORT, A FOLLOW-UP REPORT WILL BE FILED AS APPROPRIATE. MISSING INFORMATION FROM THIS REPORT IS IDENTIFIED AS BLANK; THIS INFORMATION WAS NOT PROVIDED IN THE REPORTED EVENT OR AVAILABLE AT THE TIME OF REPORT SUBMISSION. THE MANUFACTURER WILL SUBMIT A SUPPLEMENTAL REPORT IF NEW FACTS ARISE WHICH MATERIALLY ALTER INFORMATION SUBMITTED IN A PREVIOUS MDR REPORT. ADDITIONAL INFORMATION WILL BE SUBMITTED WITHIN 30 DAYS OF RECEIPT.

Additional Manufacturer Narrative · 0

MANUFACTURER¿S REF. NO: (B)(4) THE PURPOSE OF THIS MDR SUBMISSION IS TO REPORT THAT THE PRODUCT ANALYSIS LAB RECEIVED THE COMPLAINT DEVICE ON 26-FEB-2025. A SUPPLEMENTAL 3500A REPORT WILL BE SUBMITTED ONCE THE PRODUCT INVESTIGATION HAS BEEN COMPLETED. THE MANUFACTURER WILL SUBMIT A SUPPLEMENTAL REPORT IF NEW FACTS ARISE WHICH MATERIALLY ALTER INFORMATION SUBMITTED IN A PREVIOUS MDR REPORT. ADDITIONAL INFORMATION WILL BE SUBMITTED WITHIN 30 DAYS OF RECEIPT.

Additional Manufacturer Narrative · 0

MANUFACTURER¿S REF. NO:(B)(4). THE PURPOSE OF THIS MDR SUBMISSION IS TO REPORT THE INVESTIGATIONAL FINDING OF THE RETURNED DEVICE. THE COMPLAINT PRODUCT WAS RETURNED AND RECEIVED FOR EVALUATION AND ANALYSIS. THE INVESTIGATION IS DOCUMENTED BELOW. INVESTIGATION SUMMARY: A NON-STERILE 4.5MM X 28MM NO DISTAL TIP ENTERPRISE® VASCULAR RECONSTRUCTION DEVICE WAS RECEIVED CONTAINED IN THE DECONTAMINATION POUCH. VISUAL INSPECTION WAS PERFORMED. AS DESCRIBED IN THE EVENT, THE STENT COMPONENT WAS DETACHED INSIDE THE CONCOMITANT MICROCATHETER. SEVERAL KINKS WERE FOUND ALONG THE DISTAL END OF THE POSITIONING COIL OF THE DELIVERY WIRE. MICROSCOPIC INSPECTION WAS PERFORMED, AND AS A RESULT OF THE KINKS ON THE POSITIONING COIL OF THE DELIVERY WIRE, THE LOOPS WERE FOUND SLIGHTLY STRETCHED. THE STENT WAS ALSO INSPECTED UNDER THE MICROSCOPE, AND IT WAS OBSERVED TO BE IN GOOD CONDITION; THERE WAS NO STRUCTURAL DAMAGE (I.E., NO BROKEN STRUTS, NO KINKS). IT WAS ALSO NOTED TO FULLY EXPANDED WITH BOTH ENDS COMPLETELY FLARED. THE DISTAL END OF THE DELIVERY WIRE UNDERWENT DIMENSIONAL ANALYSIS. ALL MEASUREMENTS WERE FOUND TO BE WITHIN SPECIFICATIONS, INCLUDING THOSE SPECIFICATIONS THAT CONTROL THE ATTACHMENT AND DELIVERY OF THE STENT. THEREFORE, DEVICE FAILURE IS NOT SUSPECTED TO BE A CONTRIBUTING FACTOR. ALTHOUGH IN ORDER TO PERFORM A FUNCTIONAL TEST, THE STENT MUST BE ATTACHED TO THE DELIVERY WIRE AND INSIDE THE INTRODUCER, THE ISSUE REPORTED REGARDING THE ENTERPRISE BEING IMPEDED IS CONFIRMED BASED ON THE DAMAGED CONDITIONS FOUND ON THE DISTAL END OF THE DELIVERY WIRE. IT IS SUGGESTED THAT EXCESSIVE FORCE COULD HAD BEEN INADVERTENTLY APPLIED DURING THE SEVERAL ATTEMPTS TO ADVANCE THE STENT THROUGH THE MICROCATHETER, RESULTING IN THE KINKED AND STRETCHED CONDITIONS OF THE DELIVERY WIRE AND DETACHED CONDITION OF THE STENT. IT IS POSSIBLE THAT CLINICAL AND PROCEDURAL FACTORS, INCLUDING DEVICE MANIPULATION AND OPERATOR'S TECHNIQUE, MAY HAVE CONTRIBUTED TO THE REPORTED FAILURE. AT THIS TIME, THERE IS NO EVIDENCE TO SUPPORT THAT THE ISSUE REPORTED IN THE COMPLAINT IS A RESULT OF A DEFECT INHERENTLY RELATED TO THE DEVICE. THE ISSUE REPORTED REGARDING THE STENT COMPONENT BEING KINKED / BENT IS NOT CONFIRMED, AS NO DAMAGES WERE NOTED ON THE DETACHED STENT. LAKE REGION MEDICAL PERFORMED A REVIEW OF THE DEVICE HISTORY RECORDS RELATIVE TO THE MANUFACTURING, INSPECTION, AND PACKAGING OF THE LOT 9061786. THE HISTORY RECORD INDICATES THIS PRODUCT WAS FINAL INSPECTION TESTED AT LAKE REGION MEDICAL AND WAS DETERMINED TO BE ACCEPTABLE. AS PART OF JOHNSON & JOHNSON MEDTECH QUALITY PROCESS, ALL DEVICES ARE MANUFACTURED, INSPECTED, AND RELEASED TO APPROVED SPECIFICATIONS. IT SHOULD BE NOTED THAT PRODUCT FAILURE COULD BE CAUSED BY MULTIPLE FACTORS. THE INSTRUCTIONS FOR USE (IFU) DOES CONTAIN THE FOLLOWING RECOMMENDATIONS: IF RESISTANCE IS MET DURING MANIPULATION, DETERMINE THE CAUSE OF RESISTANCE BEFORE PROCEEDING. DO NOT APPLY UNDUE FORCE IF RESISTANCE IS ENCOUNTERED AT ANY POINT DURING STENT MANIPULATION. WITHDRAW THE UNIT AND ADVANCE A NEW ONE. BASED ON THE MANUFACTURING DOCUMENTATION REVIEW, THERE IS NO INDICATION THAT THE EVENT IS RELATED TO THE DEVICE MANUFACTURING PROCESS. AS PART OF THE POST MARKET SURVEILLANCE PROGRAM, INFORMATION FROM THIS COMPLAINT IS TRENDED FOR STATISTICAL SIGNALS AND CORRECTIVE / PREVENTIVE ACTION MAY BE TRIGGERED AT A LATER TIME. SINCE THERE WAS NO EVIDENCE TO SUGGEST THE EVENT WAS RELATED TO A MANUFACTURING OR DESIGN ISSUE, NO CORRECTIVE ACTIONS WILL BE TAKEN AT THIS TIME. UPDATED SECTIONS. THE MANUFACTURER WILL SUBMIT A SUPPLEMENTAL REPORT IF NEW FACTS ARISE WHICH MATERIALLY ALTER INFORMATION SUBMITTED IN A PREVIOUS MDR REPORT. ADDITIONAL INFORMATION WILL BE SUBMITTED WITHIN 30 DAYS OF RECEIPT.

Description of Event or Problem · 0

THE HEALTHCARE PROFESSIONAL REPORTED THAT DURING AN ENDOVASCULAR EMBOLIZATION PROCEDURE, THE 4.5MM X 28MM NO DISTAL TIP ENTERPRISE® VASCULAR RECONSTRUCTION DEVICE (VRD) (ENC452800 / 9061786) WAS IMPEDED IN THE PROXIMAL SECTION OF THE CONCOMITANT 150CM X 5CM PROWLER SELECT PLUS MICROCATHETER (606S255X / 31359808) AND COULD NOT ADVANCE FURTHER. THE PHYSICIAN ATTEMPTED TO RETRACT THE STENT AND OBSERVED THAT THE STENT WAS DETACHED FROM THE DELIVERY WIRE IN THE MICROCATHETER. THE PHYSICIAN REMOVED THE MICROCATHETER AND STENT FROM THE PATIENT AND REPLACED BOTH DEVICES TO COMPLETE THE PROCEDURE, WHICH WAS REPORTEDLY PROLONGED BY APPROXIMATELY 15 MINUTES. THERE WAS NO REPORT OF ANY NEGATIVE PATIENT IMPACT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
499684 CERENOVUS ENTERPRISE INTRACRANIAL NEUROVASCULAR STENT NJE MEDOS INTERNATIONAL SARL 9061786 10878528008894

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown PROWLER SELECT PLUS 150/5CM