FDA Adverse Event Malfunction Summary report: N

SYNCHROMED II

MDR report key: 2113830 · Received May 17, 2011

Report

Report Number
3004209178-2011-03603
Event Type
Malfunction
Date Received
May 17, 2011
Date of Event
January 6, 2009
Report Date
April 27, 2011
Manufacturer
MDT PUERTO RICO OPERATIONS CO., JUNCOS
Product Code
LKK
PMA / PMN Number
P860004
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MN, US
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

(B)(4). ANALYSIS OF THE PUMP (MODEL: 8637, (B)(4)) REVEALED A HYBRID ANOMALY. THE PUMP WAS RECEIVED IN AN UNOPENED BOX. A CORRUPTED CALIBRATION CONSTANT, AND PUMP MEMORY ERROR WAS INDICATED. THE MMC ANALYSIS DID NOT SHOW ANY ANOMALIES. MANY TESTS WERE CONDUCTED WHICH INVOLVED RESETTING THE CRITICAL DATA AND THE CALIBRATION CONSTANT CORRUPTED AT A SPECIFIC VOLTAGE OF 3.142. BATTERY VOLTAGE INITIALLY WAS 3.06 VOLTS. TESTING AT THESE VOLTAGES REVEALED NO CORRUPTION HAD OCCURRED. MMC TESTING, CONDUCTED TWICE, DID NOT SHOW ANY ANOMALIES. M1 AND M2 MOTOR RESISTANCE TO CASE TESTING DETERMINED THE FOLLOWING RESULTS: "OPEN." MOTOR COIL RESISTANCE WAS 492 OHMS. THE MOTOR COIL PASSED TESTING. THE PUMP PASSED PATENCY TESTING. THE PUMP ALSO PASSED A VISUAL INSPECTION. THE RESERVOIR WAS FOUND TO CONTAIN 15.0 MLS OF PF NORMAL SALINE.

Description of Event or Problem · 1

ON (B)(6) 2009, IT WAS INITIALLY REPORTED THAT A PUMP MEMORY ALARM OCCURRED. WHEN A MANUFACTURER REPRESENTATIVE TRIED READING THE PUMP, A "TELEMETRY CANCELLED" MESSAGE FOLLOWED BY A "PUMP MEMORY ERROR, CALIBRATION CONSTANT CORRUPTED" MESSAGE OCCURRED. ON (B)(6) 2009, IT WAS FURTHER REPORTED THAT A "PUMP MEMORY ERROR" OCCURRED UPON INTERROGATION OUT OF THE BOX. THERE WAS AN OPTION TO CANCEL TELEMETRY, AND IT THEN SHOWED "CALIBRATION CONSTANT CORRUPTED." THIS HAD OCCURRED IN THE OPERATING ROOM LAST NIGHT. THE PUMP WAS NOT IMPLANTED IN THE PATIENT, AS AN ALTERNATE WAS USED. THE PUMP WAS RETURNED TO THE MANUFACTURER.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 SYNCHROMED II LKK MDT PUERTO RICO OPERATIONS CO., JUNCOS 8637 NA

Patients

Seq Age Sex Outcome Treatment
1 54 YR PROGRAMMER: MODEL 8840, LOT# UNKNOWN| EXPLANTED:| EXPLANTED:| IMPLANTED:| IMPLANTED:| CATHETER: MODEL CATHETER, LOT# UNKNOWN