FDA Adverse Event Malfunction Summary report: N

MED-EL COMBI 40+ COCHLEAR IMPLANT SYSTEM

MDR report key: 2113813 · Received May 13, 2011

Report

Report Number
9710014-2011-00156
Event Type
Malfunction
Date Received
May 13, 2011
Date of Event
January 1, 2011
Report Date
May 10, 2011
Manufacturer
MED-EL ELEKTROMEDIZINISCHE GERAETE, GMBH
Product Code
MCM
PMA / PMN Number
P000025
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE DEVICE HAS NOT BEEN EXPLANTED. IF IT SHOULD BE EXPLANTED, IT IS TO BE RETURNED TO THE MANUFACTURER FOR EVALUATION. WHEN AVAILABLE, A DEVICE FAILURE ANALYSIS WILL BE SUBMITTED AS A FOLLOW UP REPORT.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE PATIENT IS ONLY RECEIVING THE SENSATION OF STIMULATION RATHER THAN HEARING SOUND. TESTING SHOWED THAT THE DEVICE HAS 2 ELECTRODE CHANNELS WITH HI AND 4 ELECTRODE CHANNELS WITH SHORT CIRCUITS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 MED-EL COMBI 40+ COCHLEAR IMPLANT SYSTEM COCHLEAR IMPLANT MCM MED-EL ELEKTROMEDIZINISCHE GERAETE, GMBH SONATA

Patients

Seq Age Sex Outcome Treatment
1 50 YR