PEN NDL 31GA 8MM 14BAG 700CAS JP
Report
- Report Number
- 3023359743-2025-00042
- Event Type
- Malfunction
- Date Received
- January 13, 2025
- Date of Event
- December 20, 2024
- Report Date
- March 4, 2025
- Manufacturer
- EMBECTA PENEL LIMITED - DUN LAOGHAIRE, IRELAND
- Product Code
- FMI
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- US
- Reporter Occupation
- OTHER
- Health Professional
- N
Narratives
3 PEN NEEDLES IMPACTED FROM LOT # 4002159.
ADDITIONAL INFORMATION PROVIDED TO SECTIONS B4, G6, H2, H3, AND H11. CORRECTIONS MADE TO SECTION H6 (TYPE OF INVESTIGATION & INVESTIGATION CONCLUSIONS). INVESTIGATION SUMMARY: NO SAMPLES (INCLUDING PHOTOS) WERE RETURNED THEREFORE THE COMPLAINT COULD NOT BE CONFIRMED AND THE ROOT CAUSE IS UNDETERMINED. THIS IS THE 1ST COMPLAINT FOR THE REPORTED LOT NUMBER. A REVIEW OF THE MANUFACTURING RECORDS WAS PERFORMED, AND NO NON-CONFORMANCES WERE RAISED IN ASSOCIATION WITH THIS TYPE OF EVENT FOR THIS LOT. COMPLAINTS RECEIVED FOR THIS DEVICE AND REPORTED CONDITION WILL CONTINUE TO BE TRACKED AND TRENDED. IF SAMPLES ARE RECEIVED IN THE FUTURE THE COMPLAINT WILL BE REOPENED FOR FURTHER INVESTIGATION. BASED ON THE ABOVE, NO ADDITIONAL INVESTIGATION AND NO CORRECTIVE/PREVENTATIVE ACTION (CAPA) OR SITUATIONAL ANALYSIS (SA) IS REQUIRED AT THIS TIME.
WHEN DID IT OCCUR? : DURING USE. NEEDLE INJURY: NO. OTHER TREATMENT: NO WERE THE RETURNED ITEMS USED? : YES. IF USED, WAS ANYTHING ADHERED TO IT? : NO. RETURNED QUANTITY: (B)(4). DETAILS OF THE EVENT (TIMELINE, HISTORY, ETC.): PRODUCT NAME: BD MICROFINE PLUS 31GX8MM CATALOG NUMBER: 320102. LOT NUMBER: 4002159. QUANTITY: (B)(4). PRODUCT 320102. FROM A CLIENT VETERINARY HOSPITAL (XXXXX), "WE WERE CONTACTED BECAUSE THREE NEEDLES DID NOT COME OUT WHEN TESTING THE MEDICINE.".
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 516600 | PEN NDL 31GA 8MM 14BAG 700CAS JP | NEEDLE, HYPODERMIC, SINGLE LUMEN | FMI | EMBECTA PENEL LIMITED - DUN LAOGHAIRE, IRELAND | 320102 | 4002159 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Prefer Not To Disclose |