30K FSI-SLI-FG-10S INS,PKD
Report
- Report Number
- 2424472-2025-00200
- Event Type
- Malfunction
- Date Received
- January 13, 2025
- Report Date
- March 28, 2025
- Manufacturer
- DENTSPLY LLC
- Product Code
- ELC
- UDI-DI
- D003820051
- PMA / PMN Number
- K052334/K970
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA
- Reporter Occupation
- DENTIST
- Health Professional
- Yes
Narratives
SINCE AN OVERHEATING INSERT COULD NECESSITATE MEDICAL/SURGICAL INTERVENTION TO PRECLUDE PERMANENT DAMAGE TO A BODY STRUCTURE OR PERMANENT IMPAIRMENT OF A BODY FUNCTION, THIS MALFUNCTION WOULD BE LIKELY TO CAUSE/CONTRIBUTE TO A SERIOUS INJURY SHOULD IT RECUR. AS SUCH, THIS EVENT MEETS THE CRITERIA FOR REPORTABILITY PER 21 CFR PART 803. THE DEVICE IS AVAILABLE FOR EVALUATION, THOUGH HAS NOT BEEN RETURNED AS OF THIS REPORT. EVALUATION RESULTS WILL BE SUBMITTED AS THEY BECOME AVAILABLE.
RETURN QTY. 1, 23300, 30K FSI-SLI-10S-KJ, (B)(6) BUL PER MANUFACTURE DATE: 30K TEST METHOD / GP005-WI02. INDUCTANCE: GAUGE ID# (B)(4) DUE: (B)(6) 2026 RESULT 3.14. MEETS SPEC DOES NOT MEET SPEC N/A. TIP CONDITION: - MEETS SPEC DOES NOT MEET SPEC N/A. STACK CONDITION MEETS SPEC DOES NOT MEET SPEC N/A. TESTED ON: G130 GAGE ID# (B)(6) DUE DATE: (B)(6) 2025 SET PRESSURE: 35 PSI WATER FLOW: OUTPUT RATE: 35 PSI - MEETS SPEC DOES NOT MEET SPEC N/A SPRAY: - MEETS SPEC DOES NOT MEET SPEC N/A WATER LEAK: HP SEALING RING PROXIMAL RING DISTAL RING SEAM LEAK N/A VIBRATION: - MEETS SPEC DOES NOT MEET SPEC N/A. GRIP CONDITION: MEET DOES NOT MEET SPEC N/A. OTHER VISUAL OBSERVATION: INSERT IS GOOD, NO FAULT FOUND. INSERT WAS TESTED ON A DIGITAL THERMOMETER I.D (B)(4). DUE DATE: (B)(6) 2025. THE TEMPERATURE WAS 88.7 °F, NO FAULT FOUND. THE SPECIFICATIONS STATE THE TEMPERATURE IS NOT TO EXCEED 118.4°F. (PER (B)(4). ALLEGED EVENT: CONFIRMED - NOT CONFIRMED.
IN THIS EVENT IT IS REPORTED THAT A 30K FSI-SLI-FG-10S INS,PKD ALLEGEDLY THE INSERT HEATS UP TOO MUCH. NO INJURY.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 462089 | 30K FSI-SLI-FG-10S INS,PKD | SCALER, ULTRASONIC | ELC | DENTSPLY LLC | UNK | D003820051 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Unknown |