ARCHITECT C4000 ANALYZER
Report
- Report Number
- 1628664-2011-00260
- Event Type
- Malfunction
- Date Received
- June 3, 2011
- Report Date
- June 1, 2011
- Manufacturer
- ABBOTT MANUFACTURING, INC.
- Product Code
- JJE
- PMA / PMN Number
- K980367
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- GH
- Reporter Occupation
- OTHER
Narratives
(B)(4). CONCOMITANT MEDICAL PRODUCTS: ARCHITECT ICT PRE-AMP BOARD PHOTO COUPLER LOTS 0834 AND 0840. EVALUATION, CONCLUSION: (B)(4) NEW WASHING PROCESS IMPACTED THE LONG TERM RELIABILITY OF THE PHOTO COUPLERS WHICH DEGRADE SLOWLY OVER TIME. THE CAUSE OF THE ARCHITECT ANALYZER PHOTO COUPLER DEGRADATION ISSUE WAS DUE TO A NEW MANUFACTURING WASHING PROCESS IMPLEMENTED BY THE SUPPLIER. ABBOTT ISSUED A PRODUCT CORRECTION LETTER TO ALL CUSTOMERS WITH AFFECTED INSTRUMENT SERIAL NUMBERS AND IMPLEMENTED A MANDATORY TECHNICAL SERVICE BULLETIN TO INSPECT ICT PRE AMP BOARDS AND REPLACE IF NEEDED.
ABBOTT FIELD SERVICE INSPECTED THE ARCHITECT ANALYZER AT THE CUSTOMER FACILITY PER (B)(4) AND REPLACED THE ARCHITECT ICT PRE-AMP BOARD AS REQUIRED. THERE WAS NO IMPACT TO PATIENT MANAGEMENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ARCHITECT C4000 ANALYZER | AUTOMATED CHEMISTRY ANALYZER | JJE | ABBOTT MANUFACTURING, INC. | C4000 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | ARCHITECT ICT PRE-AMP BOARD PHOTO COUPLER |