ASR ACETABULAR CUPS 58
Report
- Report Number
- 1818910-2011-08903
- Event Type
- Injury
- Date Received
- May 20, 2011
- Date of Event
- December 3, 2013
- Report Date
- January 28, 2014
- Manufacturer
- DEPUY INTERNATIONAL, LTD.
- Product Code
- KWA
- PMA / PMN Number
- K040627
- Removal / Correction Number
- Z-1749/1816-2011
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- LA, US
- Reporter Occupation
- OTHER
Narratives
DEPUY STILL CONSIDERS THIS INVESTIGATION CLOSED.
THE ASR PLATFORM WAS VOLUNTARILY RECALLED FROM THE MARKET IN AUGUST 2010, AND THE ASR PRODUCT CODES ARE NOW CONSIDERED INACTIVE. FURTHER INVESTIGATION OF THIS INDIVIDUAL INCIDENT WILL NOT BE UNDERTAKEN, AS THERE IS AN ONGOING INVESTIGATION REGARDING THE ROOT CAUSE(S) AND/OR CORRECTIVE ACTIONS. REF. WWCAPA (B)(4). DEPUY CONSIDERS THE INVESTIGATION CLOSED AT THIS TIME. SHOULD THE PRODUCT AND/OR ADDITIONAL INFORMATION BE RECEIVED, THE INVESTIGATION WILL BE RE-OPENED.
DEPUY STILL CONSIDERS THIS INVESTIGATION CLOSED.
ADDITIONAL NARRATIVE: IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE. DEPUY STILL CONSIDERS THIS INVESTIGATION CLOSED.
THIS COMPLAINT IS STILL UNDER INVESTIGATION. DEPUY WILL NOTIFY THE FDA OF THE RESULTS OF THIS INVESTIGATION ONCE IT HAS BEEN COMPLETED.
LITIGATION PAPERS ALLEGE SINCE SURGERY, PT HAS SUFFERED SYMPTOMS INCLUDING, BUT NOT LIMITED TO PAIN, SWELLING, SORENESS, DIFFICULTY WALKING, AND DECREASED MOBILITY.
LITIGATION PAPERS ALLEGE SINCE SURGERY, PATIENT HAS SUFFERED SYMPTOMS INCLUDING, BUT NOT LIMITED TO PAIN, SWELLING, SORENESS, DIFFICULTY WALKING, AND DECREASED MOBILITY. UPDATE: (B)(6) 2012 - PLAINTIFFS PRELIMINARY DISCLOSURE FORM WAS RECEIVED, WHICH IDENTIFIED PART/LOT INFORMATION. THE COMPLAINT AND ASSOCIATED MDRS WERE UPDATED. THERE WAS NO NEW INFORMATION THAT WOULD CHANGE THE OUTCOME OF THE INVESTIGATION.
**UPDATE**(B)(4) 2013 - SALES REP REPORTED REVISION SURGERY. PATIENT WAS REVISED TO ADDRESS INSTABILITY. ELECTIVE SURGERY. THERE IS NO NEW ADDITIONAL INFORMATION THAT WOULD AFFECT THE INVESTIGATION. THE COMPLAINT WAS UPDATED ON: (B)(4) 2013.
UPDATE: (B)(4) 2014 - MEDICAL RECORDS RECEIVED. PATIENT WAS REVISED TO ADDRESS FLUID, CAPSULITIS, THICK FIBROUS TISSUE, METAL REACTION, CORROSION AT THE TRUNNION, INFLAMED TISSUE, METAL DEBRIS, AND A CYST. THE STEM AND SLEEVE HAVE BEEN ADDED TO THE COMPLAINT. STEM REMAINED IN SITU. THIS COMPLAINT WAS UPDATED ON: (B)(4) 2014.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ASR ACETABULAR CUPS 58 | TOTAL HIP REPLACEMENT | KWA | DEPUY INTERNATIONAL, LTD. | NA | 2352762 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Required Intervention |