FDA Adverse Event
Malfunction
Summary report: N
GYNECARE X-TRACT MORCELLATOR
MDR report key: 2113218
·
Received June 3, 2011
Report
- Report Number
- 2210968-2011-00705
- Event Type
- Malfunction
- Date Received
- June 3, 2011
- Date of Event
- May 11, 2011
- Report Date
- May 12, 2011
- Manufacturer
- ETHICON
- Product Code
- HET
- PMA / PMN Number
- K993801
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MI, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
(B)(4). DAMAGE TO DRIVE CONNECTOR OR COUPLING. CONCLUSION: THE PRODUCT UPON WHICH THIS MEDWATCH IS BASED IS ANTICIPATED. ONCE THE PRODUCT IS RECEIVED, ANY FURTHER INFORMATION DERIVED FROM THE EVALUATION WILL BE SUBMITTED IN A SUPPLEMENTAL 3500A FORM.
Additional Manufacturer Narrative · 1
(B)(4). CONCLUSION: THE ACTUAL DEVICE INVOLVED IN THIS EVENT WAS RETURNED FOR EVALUATION. THE DEVICE WAS VISUALLY AND FUNCTIONALLY EVALUATED. THE DEVICE WAS FOUND TO HAVE A BROKEN CPC CONNECTOR.
Description of Event or Problem · 1
IT WAS REPORTED THAT A PATIENT UNDERWENT A GYNECOLOGICAL PROCEDURE ON (B)(6) 2011. DURING THE PROCEDURE, THE CPC CONNECTOR BROKE. THE PROCEDURE WAS COMPLETED WITH THE SAME DEVICE WITH NO ADVERSE PATIENT CONSEQUENCES.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | GYNECARE X-TRACT MORCELLATOR | LAPAROSCOPIC MORCELLATOR | HET | ETHICON | NA | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |