FDA Adverse Event Malfunction Summary report: N

ANIMAS INSULIN INFUSION PUMP

MDR report key: 2113073 · Received June 3, 2011

Report

Report Number
2531779-2011-03900
Event Type
Malfunction
Date Received
June 3, 2011
Date of Event
May 10, 2011
Report Date
May 10, 2011
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K042873
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NY, US
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

THE PUMP HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON (B)(6) 2011 WITH THE FOLLOWING FINDINGS: EVALUATION FOUND THAT THE PUMP POWERS UP PROPERLY. POWER FLEX, CIRCUIT BOARD, AND TERMINAL CONNECTIONS WERE FOUND TO BE INTACT. THE PUMP WAS EXERCISED FOR 24 HOURS WITH NO RE-BOOTS OCCURRING. A REVIEW OF PUMP HISTORY INDICATED THAT PUMP HAD REBOOTED WHICH WAS NOT DUPLICATED DURING TESTING. THE CONTRAST BUTTON WAS FOUND TO NOT SPRING BACK AS NORMAL. THE CONTRAST BUTTON HAS NO EFFECT ON THE PUMP'S INSULIN DELIVERY FUNCTION. THE KEYPAD WAS REMOVED AND ADHESIVE WAS OBSERVED UNDER ALL OF THE KEYPAD BUTTON CONTACTS.

Description of Event or Problem · 1

THE PATIENT REPORTED THAT THE PUMP KEEPS REBOOTING AND GOING BACK TO THE VERIFY SCREEN. SHE STATED THAT SHE REPLACED THE BATTERY. THE PATIENT ALSO STATED THAT KEYPAD BUTTONS ARE HARD TO PRESS AND NOT RESPONDING APPROPRIATELY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ANIMAS INSULIN INFUSION PUMP INSULIN INFUSION PUMP LZG ANIMAS CORPORATION ANM 2020 INSULIN INFUSION PUMP

Patients

Seq Age Sex Outcome Treatment
1 21 YR