FDA Adverse Event Malfunction Summary report: N

ESPRIT VENTILATOR

MDR report key: 2113044 · Received June 3, 2011

Report

Report Number
2031642-2011-00174
Event Type
Malfunction
Date Received
June 3, 2011
Date of Event
April 30, 2011
Report Date
May 9, 2011
Manufacturer
RESPIRONICS CALIFORNIA, INC.
Product Code
CBK
PMA / PMN Number
K981072
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
RI, US
Reporter Occupation
RESPIRATORY THERAPIST

Narratives

Description of Event or Problem · 1

THE CUSTOMER REPORTED THE VENTILATOR ALARMED AND RESTARTED WHILE IN USE ON A PATIENT. THE CUSTOMER REPORTED THERE WAS NO PATIENT HARM. THE MANUFACTURER'S SERVICE TECHNICIAN WAS UNABLE TO DUPLICATE THE CUSTOMER REPORTED PROBLEM. THE SERVICE TECHNICIAN VERIFIED A DIAGNOSTIC CODE IN THE VENTILATOR LOG HISTORY THAT INDICATED A VENTILATOR RESTART. THE SERVICE TECHNICIAN REPLACED THE POWER SWITCH TO ADDRESS THE FINDING. PERFORMANCE VERIFICATION TESTING WAS COMPLETED AND TESTS PASSED PER OPERATING SPECIFICATIONS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ESPRIT VENTILATOR VENTILATOR, CONTINUOUS CBK RESPIRONICS CALIFORNIA, INC. V1000 NA

Patients

Seq Age Sex Outcome Treatment
1