FDA Adverse Event
Malfunction
Summary report: N
CENTRA BED
MDR report key: 2112917
·
Received May 25, 2011
Report
- Report Number
- 1824206-2011-02885
- Event Type
- Malfunction
- Date Received
- May 25, 2011
- Date of Event
- May 4, 2011
- Report Date
- May 4, 2011
- Manufacturer
- HILL-ROM, INC.
- Product Code
- FNL
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- OH, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
THE TECHNICIAN FOUND THE BRAKE CABLE WAS PINCHED BETWEEN THE STIFFENER PLATE AND BEARING, AND THE BUSHINGS WERE WORN WITH GOUGES AND THE TOP BLOCK BEARING POST WAS BROKEN. THE TECHNICIAN REBUILT THE BRAKE BLOCK MECHANISM AND ADJUSTED THE BRAKES TO REPAIR THE BED.
Description of Event or Problem · 1
INFORMATION RECEIVED INDICATES THE BRAKE WILL SET BUT DOES NOT HOLD.