FDA Adverse Event Malfunction Summary report: N

CENTRA BED

MDR report key: 2112917 · Received May 25, 2011

Report

Report Number
1824206-2011-02885
Event Type
Malfunction
Date Received
May 25, 2011
Date of Event
May 4, 2011
Report Date
May 4, 2011
Manufacturer
HILL-ROM, INC.
Product Code
FNL
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
OH, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE TECHNICIAN FOUND THE BRAKE CABLE WAS PINCHED BETWEEN THE STIFFENER PLATE AND BEARING, AND THE BUSHINGS WERE WORN WITH GOUGES AND THE TOP BLOCK BEARING POST WAS BROKEN. THE TECHNICIAN REBUILT THE BRAKE BLOCK MECHANISM AND ADJUSTED THE BRAKES TO REPAIR THE BED.

Description of Event or Problem · 1

INFORMATION RECEIVED INDICATES THE BRAKE WILL SET BUT DOES NOT HOLD.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 CENTRA BED AC-POWERED ADJUSTABLE HOSPITAL BED FNL HILL-ROM, INC. 850

Patients

Seq Age Sex Outcome Treatment
1