FDA Adverse Event Malfunction Summary report: N

ADVANTA BED

MDR report key: 2112915 · Received May 25, 2011

Report

Report Number
1824206-2011-02887
Event Type
Malfunction
Date Received
May 25, 2011
Date of Event
May 4, 2011
Report Date
May 4, 2011
Manufacturer
HILL-ROM, INC.
Product Code
FNL
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NY, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE TECHNICIAN REPLACED THE POWER CORD PLUG TO REPAIR THE BED.

Description of Event or Problem · 1

INFORMATION RECEIVED INDICATES THE GROUND RESISTANCE IS HIGH ON THE BED DUE TO A LOOSE GROUND PIN ON THE POWER CORD PLUG.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ADVANTA BED AC-POWERED ADJUSTABLE HOSPITAL BED FNL HILL-ROM, INC. 1604

Patients

Seq Age Sex Outcome Treatment
1