HANAULUX 2003 LIGHT
MDR report key: 21129
·
Received January 24, 1995
Report
- Report Number
- 2919178-1995-00002
- Event Type
- Other
- Date Received
- January 24, 1995
- Date of Event
- October 17, 1994
- Report Date
- March 17, 1995
- Manufacturer
- HERAEUS INSTRUMENTS
- Product Code
- FQP
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Distributor report
- Reporter Location
- FL, US
- Reporter Occupation
- OTHER
Narratives
Description of Event or Problem · 1
FOLLOW-UP REPORT 2/17/95DURING A FIELD SERVICE VISIT, IT WAS DETERMINED THAT THE OPERATING ROOM LIGHTHEADS WERE LONDON LIGHTS, MOUNTED ON A MODEL 22000 SERIES SPRING ARM. THE LIGHT ASSEMBLY WAS INSTALLED BY A PREVIOUS DISTRIBUTOR. THE RETAINING COLLAR WAS FOUND TO BE IN THE UP POSITION, WHICH ALLOWED THE CLIP TO COME OUT AND THE LIGHT HEAD TO FALL. THE RETAINING CLIP WAS NOT RETREIVED.THE HOSPITAL REPORTED THAT THE PT WAS UNINJURED, HOWEVER THE NURSE WAS HIT ON THE FACE IN AN AREA WHERE SHE HAD BEEN PREVIOUSLY INJURED DURING AN UNRELATED INCIDENT. NO FURTHER DETAILS OF THE EXTENT OF THE NURSES INJURY WERE MADE AVAILABLE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | HANAULUX 2003 LIGHT | OPERATING ROOM LIGHT | FQP | HERAEUS INSTRUMENTS | 2003 | 0288007 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | * | Other |