FDA Adverse Event Other Summary report: N

HANAULUX 2003 LIGHT

MDR report key: 21129 · Received January 24, 1995

Report

Report Number
2919178-1995-00002
Event Type
Other
Date Received
January 24, 1995
Date of Event
October 17, 1994
Report Date
March 17, 1995
Manufacturer
HERAEUS INSTRUMENTS
Product Code
FQP
Adverse Event
Yes
Product Problem
Yes
Report Source
Distributor report
Reporter Location
FL, US
Reporter Occupation
OTHER

Narratives

Description of Event or Problem · 1

FOLLOW-UP REPORT 2/17/95DURING A FIELD SERVICE VISIT, IT WAS DETERMINED THAT THE OPERATING ROOM LIGHTHEADS WERE LONDON LIGHTS, MOUNTED ON A MODEL 22000 SERIES SPRING ARM. THE LIGHT ASSEMBLY WAS INSTALLED BY A PREVIOUS DISTRIBUTOR. THE RETAINING COLLAR WAS FOUND TO BE IN THE UP POSITION, WHICH ALLOWED THE CLIP TO COME OUT AND THE LIGHT HEAD TO FALL. THE RETAINING CLIP WAS NOT RETREIVED.THE HOSPITAL REPORTED THAT THE PT WAS UNINJURED, HOWEVER THE NURSE WAS HIT ON THE FACE IN AN AREA WHERE SHE HAD BEEN PREVIOUSLY INJURED DURING AN UNRELATED INCIDENT. NO FURTHER DETAILS OF THE EXTENT OF THE NURSES INJURY WERE MADE AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 HANAULUX 2003 LIGHT OPERATING ROOM LIGHT FQP HERAEUS INSTRUMENTS 2003 0288007

Patients

Seq Age Sex Outcome Treatment
1 * Other