PFC SIGMARP STB TB IN 4 10.0
Report
- Report Number
- 1818910-2011-08846
- Event Type
- Injury
- Date Received
- May 18, 2011
- Date of Event
- April 19, 2011
- Report Date
- April 19, 2011
- Manufacturer
- DEPUY-CORK, A DIVISION OF DEPUY ORTHOPAEDICS, INC.
- Product Code
- NJL
- PMA / PMN Number
- P830055/S074
- Removal / Correction Number
- NA
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FL, US
- Reporter Occupation
- OTHER
Narratives
THE PRODUCT ASSOCIATED WITH THIS REPORTED EVENT WAS NOT RETURNED FOR EXAMINATION. A SEARCH OF THE COMPLAINT DATABASE DID NOT SHOW ANY ADDITIONAL REPORTS AGAINST THE LOT CODE. THE INVESTIGATION COULD NOT DRAW ANY CONCLUSIONS ABOUT THE REPORTED EVENT; HOWEVER, PROVIDED INFORMATION SUGGESTS POOR DEVICE POSITIONING WAS A CONTRIBUTING FACTOR. ADDITIONAL PROVIDED INFORMATION STATED THE PRODUCT WAS NOT SUSPECTED OF FAILING TO MEET SPECIFICATIONS. NO EVIDENCE WAS FOUND SUGGESTING PRODUCT ERROR WAS A CONTRIBUTING FACTOR AND THE NEED FOR CORRECTIVE ACTION IS NOT INDICATED. DEPUY CONSIDERS THE INVESTIGATION CLOSED AT THIS TIME. SHOULD THE PRODUCT AND/OR ADDITIONAL INFORMATION BE RECEIVED, THE INVESTIGATION WILL BE RE-OPENED.
PATIENT WAS REVISED TO ADDRESS PAIN. THE SURGEON REMOVED THE INSERT AND FOUND BONE BEING HIT BY THE RP INSERT AS IT ROTATED. HE REMOVED THE BASEPLATE AND CONVERTED TO A FIXED-BEARING KNEE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | PFC SIGMARP STB TB IN 4 10.0 | 87NJL | NJL | DEPUY-CORK, A DIVISION OF DEPUY ORTHOPAEDICS, INC. | NA | 2073481 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 59 YR | Required Intervention |